50
Removing the Infusion Set
Peel up one end of the tape as shown in Figure 58 and
continue to peel away from the skin until the entire
infusion set has been removed.
Filling the Reservoir
1.
Remove the new reservoir from the packaging.
2.
Use an alcohol wipe to clean the insulin vial and
then attach it to the fill adapter as in Figure 59.
3.
With the insulin vial on top, slowly pull the pullrod
and draw insulin into the reservoir as in Figure 60.
4.
Tap the side of the reservoir to encourage any air
bubbles to rise to the top of the reservoir (Figure
61).
5.
Push the pullrod slowly to eject the bubbles back
into the vial, and then slowly pull to draw more
insulin into the reservoir. Repeat until the reservoir
has enough insulin and there are no air bubbles
(Figure 62).
Figure 58
Figure 59
Figure 60
Figure 61
Figure 62