24
a.
The Bolus Increment setting (Figure 19) allows
you to set the interval used when increasing or
decreasing bolus amounts.
b.
The Maximum Bolus setting allows you to set an
upper limit on the amount of insulin given in a
single bolus (Figure 20). Use the “+” and “-“ buttons
to select the value. Choose OK to save and exit or
Cancel to exit without saving.
Figure 19
Figure 20
Figure 21
Note:
This feature is used to limit the maximum
amount of bolus delivery that can be
administered by accident or misoperation.
c.
Choose the Extended Bolus setting to open a dialog
box (Figure 21) that will give you options.
Off:
Extended Bolus function is disabled. Extended
Bolus controls will be greyed out.
U:
Amount now will be shown in U (units of Insulin).
%:
Amount now will be shown as a percentage of
the entire bolus.
After you have finished choosing the Basic Bolus
settings, choose Next to enter the fourth Settings
Wizard screen, or Back to enter the previous setting
screen.
4. Advanced Bolus Settings
The fourth screen in the Settings Wizard shows the Advanced Bolus Settings
(Figure 22).