46
Note:
Cancelling a bolus delivery will cancel the total bolus (Amount Now and
Extended amount).
Note:
This feature is used to reduce the possibility of overdosage caused by
mistake. Try to set this value as the maximum bolus that you would
normally give. Setting this value too high increase the risk of overdosage
by misoperation.
Settings
From the Home Screen, the Bolus settings can be found
at “Settings - Bolus Settings” as shown in Figure 51.
1. Bolus Rate:
The normal bolus delivery rate is 3.0U/
min. Use this menu to decrease the bolus rate to
1.5U/min.
2. Bolus Increment Setting:
This is the amount that
the requested bolus changes when pressing the “+”
or “-” button.
3. Maximum Bolus Setting:
This is the maximum
bolus that the user can choose in the bolus menu.
Press the Cancel button and a confirmation window will appear (in PDA). Press OK
to confirm. You can also cancel a bolus by pressing and holding the Quick Bolus
button on the pump to cancel.
Figure 51
4. Extended Bolus Settings:
a. Off:
Extended bolus function is off, and corresponding button will be greyed
out (default)
b. U:
The Amount Now will be displayed in Units of insulin (U).
c. %:
The Amount Now will be displayed as a % of the total bolus amount.