108
IEC 60601-1-2: Table 202
Guidance and manufacturer’s declaration – electromagnetic immunity
The device is intended for use in the electromagnetic environment specified
below. The customer or the user of the device should ensure that it is used in such
an environment.
Immunity Test
IEC 60601 Test Level
Compliance Level
Electromagnetic environment –
guidance
Electrostatic discharge
(ESD) IEC 60601-4-2
±6 kV Contact
±8 kV Air
±8 kV Contact
±15 kV Air
Floors should be wood, concrete
or ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30 %.
Electrical fast transient
burst IEC 61000-4-4
±2 kV
±1 kV for input/output
lines
±2 kV
±1 kV for input/output
lines
Mains power quality should be that
of a typical commercial or hospital
environment.
Surge IEC 61000-4-5
±1 kV Differential mode
±2 kV Common mode
±1 kV Differential mode
±2 kV Common mode
Mains power quality should be that
of a typical commercial or hospital
environment.
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines IEC 61000-4-11
<5% Ut (>95 % dip in
Ut) for 0,5 cycle
40% Ut (60 % dip in Ut)
for 5 cycles
70% Ut (30 % dip in Ut)
for 25 cycles
<5% Ut (>95 % dip in
Ut) for 5 s
<5% Ut (>95 % dip in Ut)
for 0,5 cycle
40% Ut (60 % dip in Ut)
for 5 cycles
70% Ut (30 % dip in Ut)
for 25 cycles
<5% Ut (>95 % dip in Ut)
for 5 s
Mains power quality should be that
of a typical commercial or hospital
environment. If the user of the
device requires continued operation
during power mains interruptions,
it is recommended that the device
be powered from an uninterruptible
power supply or a battery.
Power frequency
(50/60 Hz) magnetic
field IEC 61000-4-8
3 A/m
400 A/m
The power frequency magnetic
field should have the characteristics
of power frequency magnetic
field level at a typical place in a
typical commercial and hospital
environment