18
TECHNICAL SPECIFICATIONS
Channel: Dual , 2 wireless units independent and individually adjustable channels, electrically insulated from one another.
Pulse Rate : As pre-programmed, in operation mode.
Pulse Width : As pre-programmed, in operation mode.
Timer : 5~60 min. selectable.
Wave Form : Symmetrical Bi-Phasic square pulse.
Max Charge per Pulse : 20.8 micro-coulombs maximum.
** All electrical specifications are ±10% at 500
Ω
load.
Input rating: 5Vdc, 1A
Power Source : Remote Control Battery 3.7V /260 mAh Lithium polymer (LiPo) battery.
Wireless unit
3.7V /
180 mAh Lithium polymer (LiPo) battery.
AC Adaptor manufacturer/model: FranMar International Inc. / FRM06-S05-z, 28.1x64.1x42 mm( LxWxH)
AC Adaptor I/P: 100-240Vac, 50-60Hz, 0.3-0.15A ; O/P: 5Vdc, 1.2A
Weight & Dimension: Remote Control 60 grams (battery included), 156.98x47x15.50 mm
Wireless Unit 30 grams (battery included) , 60x60x15.65 mm
Operating Conditions: + 50°F (10
℃
) to +104° (40
℃
), 40-90% max. Relative humidity
Transport and Storage
Conditions
: +14°F (-10
℃
) to +140° (60
℃
), 30-95% max. Relative humidity
Operation altitude: 3000m.
Operating Atmospheric pressure range: 700~1013 hPa
Transport and Storage Atmospheric pressure range: 500~1060 hPa
RF Module :
Radiations Frequency
:
2.4GHz
Frequency Band
:
2410~~2450MHz
GFSK typical
:
GFSK
(i) There are a number of technical symbols on your system, explained as follows:
Remote Label Receiver Label
This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions:
(1) this device may not cause harmful interference, and (2) this device must accept any interference received,
(2) including interference that may cause undesired operation.