2

 

Do not use this System while sleeping. 

 

Do not use if you feel numbness. 

 

Do not use this System in or close to water.  

 

Do not apply stimulation across the chest because the introduction of electrical current into the chest may cause 
rhythm disturbances to the heart, which could be lethal.  

 

Do not use the pads over or close to cancerous lesions. 

 

Use the pads only on normal, healthy, clean and dry skin. Do not use the pads on open wounds or rashes, or over 
swollen, red, infected or inflamed skin. 

 

If you have ever had back surgery, consult your Physician before using this System. 

 

Electronic monitoring equipment (such as ECG and ECG alarms) may not operate properly when electrical stimulation 
is in use. 

 

You must position the pads and operate the unit ONLY as indicated in this manual. 

 

Avoid areas in injury or restricted movement (e.g. fractures or sprains) 

 

Avoid placing the pads over metal implants. 

 

Do not use in the bath or shower, or in an environment of elevated humidity (e.g. Sauna, hydrotherapy, etc).  

 

Operation in close proximity, such as 3 feet (1 meter), to shortwave or microwave therapy equipment may produce 
instability in the device output and may shut the device off. 

 

Do not use the device in an environment where flammable or explosive fumes may exist. 

 

Patient should never operate potentially dangerous machinery such as power saws, automobiles, etc. during electrical 
stimulation. It might result in burs at the site of the stimulator electrodes and possible damage to the stimulator if you 
connect to high frequency equipment.  

 

Application of electrodes near the thorax may increase the risk of cardiac fibrillation.  

 

The system is internally powered (power form battery pack), IP 22, continuous operation.  

 

Wait before using the unit : 
 

 

 At least 6 weeks after the birth of your baby (you must consult your doctor before use). 

 

One month after an IUD contraceptive device (e.g. coil) has been fitted (you must consult your doctor before use). 

 

At least 3 months after having a caesarean section (you must consult your doctor before use). 

 

The heavy days of your period have finished, because vigorous abdominal exercise is not recommended at this time.  

 

 

Precautions 

 

 

Read User Manual before using this System for the first time. 

 

Keep this manual available whenever you use your System. 

 

The System is intended for personal use on healthy adult muscle only.  

 

The safety of using the System during pregnancy or birth has not been established. 

 

The effectiveness of the System depends greatly on a person’s individual physical condition. It may not always be 
effective for every user. 

 

The safety of neuromuscular stimulation during pregnancy has not been established. 

 Use 

caution 

when/if: 

-

 

Sensory nerve damage is present by a loss of normal skin sensation. 

-

 

Use caution prior to using this device on patients suspected of having heart disease. 

-

 

Use caution for patients with suspected or diagnosed epilepsy when using this device. 

-

 

Use caution following recent surgical procedures when muscle contraction may disrupt the healing process. 

-

 

Use caution when there is a tendency to hemorrhage, such as following acute trauma or fracture. 

-

 

A menstruating or pregnant uterus. 

-

 

Patient experiences skin irritation due to electrical stimulation or the electrical conductive medium used, remove 
the electrodes, discontinue stimulation, and consult the clinician. Irritation may be reduced by an alternative 
conductive medium or an alternative electrode placement. Isolated cases of skin irritation may occur at the site of 
electrode placement following long term application. 

 

Place electrodes in accordance with illustrations in the User Manual. 

 

This unit should not be used while driving, operating machinery or during any activity in which involuntary muscle 
contractions may place the user at undue risk of injury. 

 

Some users may experience skin irritation or hypersensitivity due to the electrical stimulation or the conductive 
medium. 

 

Keep this device out of the reach of children. If the patient is a child, make sure he/she is properly supervised during 
electrical stimulation. Strangulation resulting from baby or child entanglement in monitoring cables.  

Содержание WR-2605A OTC

Страница 1: ...FE Wireless TENS EMS Stimulator Model no WR 2605A OTC WELL LIFE HEALTHCARE LTD 1Fl No 16 Lane 454 Jungjeng Rd Yunghe District New Taipei City 23455 Taiwan Operation Manual Read Before Using IM WR2605A...

Страница 2: ...tem via USB Port STEP BY STEP OPERATION GUIDE FOR TRAINING 8 Preparing Skin for Running a Session Placement of Electrodes Turning ON OFF Remote Control Turning ON OFF Wireless Unit Communication betwe...

Страница 3: ...ONS Warnings If you are under the care of a Physician consult with your Physician before using this System The long term effects of this System are not known Do not place the pads on or close to your...

Страница 4: ...ng a caesarean section you must consult your doctor before use The heavy days of your period have finished because vigorous abdominal exercise is not recommended at this time Precautions Read User Man...

Страница 5: ...l stimulation If this occurs stop using your System and consult your Physician If skin under one of more pads feels irritated after using the stimulator for a long period of time use the stimulator fo...

Страница 6: ...your eyes and to your head and face You should stop using the device and should consult with your physician if you experience adverse reactions from the device Conditions that may affect your Well Lif...

Страница 7: ...pcs bag 4 USB Charging Cables x 2 Length 60 cm 5 AC Adaptor x1 About Remore Control 1 Power ON key 2 CH key 3 SET key 4 Power OFF key 5 ON key increase 6 OFF key decrease 7 USB charger socket About R...

Страница 8: ...ptor 2 Plug the adaptor into any standard wall outlet When the remote has finished charging the battery an icon will indicate full Charging the Receiver 1 Connect the small end of the USB cable to the...

Страница 9: ...a strong impact heavy blow etc Do not use sharp utensils or metal objects to scratch or puncture the battery Do not place the battery into a microwave oven or dryer Do not place the battery into high...

Страница 10: ...termine the placement sites for the electrodes 2 Wash the area with mild soap and water do not use alcohol Rinse and dry thoroughly 3 Trim excess body hair from the area with scissors do not shave 4 O...

Страница 11: ...o a flat surface Place the wireless unit with the contact jacks down onto the Electrode pad so that plugs and jacks are aligned Then press down until completely snapped in 2 Do not remove the clear pl...

Страница 12: ...bear in mind that the muscle contraction may cause involuntary limb movement which could cause injury to you or others Always ensure the limb is secured to prevent movement Turning On Off Receiver 1...

Страница 13: ...blinking on the Channel in use when it receives signal transmitting from the Remote Control 3 LED in Red Flashing battery power is low Receiver unit will turn on when charging the battery Orange light...

Страница 14: ...selected modality Note Make sure the identical receiver is turned on and properly placed For TENS programs When using any of the 8 programs for pain relief always start with the lowest intensity and g...

Страница 15: ...fortable tingling and pumping action action should appear to come in waves P8 for symptomatic relief and management of chronic intractable pain and relief of pain associated with arthritis Variable co...

Страница 16: ...d capillary supply It improves fatigue resistance during long duration moderate intensity exercise The exercise comprises an alternating sequence of work and rest phases lasting several seconds Increa...

Страница 17: ...The Well Life Wireless TENS EMS System is recommended for single person use at home Remote and Receivers The system device s may be wiped clean with a small amount of soapy water on a clean cloth Do n...

Страница 18: ...pt to disassemble or force open the built in battery Batteries have a finite lifetime for 12 months To maximize battery life please charge the system every 3 months when not in use If you find the bat...

Страница 19: ...harging Ensure the USB Cable is correctly into the Remote Receiver s and firmly plugged into the AC adaptor and into the electrical socket on the other side Receiver s does not pair with remote The re...

Страница 20: ...0 15A O P 5Vdc 1 2A Weight Dimension Remote Control 60 grams battery included 156 98x47x15 50 mm Wireless Unit 30 grams battery included 60x60x15 65 mm Operating Conditions 50 F 10 to 104 40 40 90 max...

Страница 21: ...quid or spray any liquids directly on power unit Stand by ii There is a label on the package explained as This symbol means used before represent as YYYY MM for year and month ELECTROMAGNETIC COMPATIB...

Страница 22: ...37 100 11 67 11 67 23 33 Declaration electromagnetic emissions and immunity for EQUIPMENT and SYSTEMS that are not LIFE SUPPORTING and are specified for use only in a shielded location The Wireless El...

Страница 23: ...ip in UT for 25 cycles 5 UT 95 dip in UT for 5 sec Mains power quality should be that of a typical commercial or hospital environment If the user of the EQUIPMENT or SYSTEM requires continued operatio...

Страница 24: ...off and on the user is encouraged to try to correct the interference by one or more of the following measures Reorient or relocate the receiving antenna Increase the separation between the equipment...

Страница 25: ...to be free from defects in workmanship and materials at the time of delivery The warranty does not apply to damage resulting from failure to follow the operating instructions accidents abuse alterati...

Отзывы: