background image

 

12

 

The Device offers 8 different preset treatment programs respectively for TENS and 6 programs for EMS modes; the 

programs differ with respect to varying pulse widths and frequencies.  

 

Choice of the appropriate mode and treatment program 

The mode you choose determines the kind of work that is imposed upon the stimulated muscles. Choose the mode that is 

appropriate to your needs or gives you the greatest pleasure.  

 

 

How to Select TENS/EMS Mode   

 
 
1.  Press the “SET”  button , the preset (default) therapy  

mode TENS/EMS will flash on the display.  
 

Use ON + or the button OFF –, if you would like to 
change the therapy mode. 

 

2.  Press the SET button again, the numeric number of 

program is then flashing. 

 
3.  Press the button ON + (to increase) or the button 

OFF – (to decrease) for choice of program of the 
selected modality. 

 

Note 

: Make sure the identical receiver is turned on and 

properly placed.   

 

  

 

 

                      

   

 

 

 

 

 

 

For 

TENS 

programs:  

When using any of the 8 programs for pain relief always start with the lowest intensity and gradually increase the  

level of intensity until you feel a “tingling” sensation. All programs are different and therefore feel differently. You may  

try all 8 programs in the beginning and choose one that feels pleasant. Never increase the intensity to a level so that it hurts, 

always stay under the point of discomfort. Start with short sessions of 5 or 10 minutes until your body gets used to the 

stimulation.  

Program 

Max. 

Phase duration 

Rate 

Function mode 

P1 80mA 

260uS 

15Hz 

Constant 

P2 80mA 

260uS 

60Hz 

Modulated 

P3 80mA 

260uS 

60Hz 

Constant 

P4 

80mA 

260~150uS 

2 ~ 60Hz 

Modulated 

P5 80mA 

260~150uS 

60Hz 

Modulated 

P6 80mA 

260uS 

<->60Hz  Modulated 

P7 80mA  260~156 

uS 

60Hz 

Modulated 

P8 

80mA 

210 uS 

2.45~245 Hz 

Cycle 

All electrical specification ±20% 
 
 
 
 
 
 

Program./mode 

Benefits 

You should feel 

Содержание WR-2605A OTC

Страница 1: ...FE Wireless TENS EMS Stimulator Model no WR 2605A OTC WELL LIFE HEALTHCARE LTD 1Fl No 16 Lane 454 Jungjeng Rd Yunghe District New Taipei City 23455 Taiwan Operation Manual Read Before Using IM WR2605A...

Страница 2: ...tem via USB Port STEP BY STEP OPERATION GUIDE FOR TRAINING 8 Preparing Skin for Running a Session Placement of Electrodes Turning ON OFF Remote Control Turning ON OFF Wireless Unit Communication betwe...

Страница 3: ...ONS Warnings If you are under the care of a Physician consult with your Physician before using this System The long term effects of this System are not known Do not place the pads on or close to your...

Страница 4: ...ng a caesarean section you must consult your doctor before use The heavy days of your period have finished because vigorous abdominal exercise is not recommended at this time Precautions Read User Man...

Страница 5: ...l stimulation If this occurs stop using your System and consult your Physician If skin under one of more pads feels irritated after using the stimulator for a long period of time use the stimulator fo...

Страница 6: ...your eyes and to your head and face You should stop using the device and should consult with your physician if you experience adverse reactions from the device Conditions that may affect your Well Lif...

Страница 7: ...pcs bag 4 USB Charging Cables x 2 Length 60 cm 5 AC Adaptor x1 About Remore Control 1 Power ON key 2 CH key 3 SET key 4 Power OFF key 5 ON key increase 6 OFF key decrease 7 USB charger socket About R...

Страница 8: ...ptor 2 Plug the adaptor into any standard wall outlet When the remote has finished charging the battery an icon will indicate full Charging the Receiver 1 Connect the small end of the USB cable to the...

Страница 9: ...a strong impact heavy blow etc Do not use sharp utensils or metal objects to scratch or puncture the battery Do not place the battery into a microwave oven or dryer Do not place the battery into high...

Страница 10: ...termine the placement sites for the electrodes 2 Wash the area with mild soap and water do not use alcohol Rinse and dry thoroughly 3 Trim excess body hair from the area with scissors do not shave 4 O...

Страница 11: ...o a flat surface Place the wireless unit with the contact jacks down onto the Electrode pad so that plugs and jacks are aligned Then press down until completely snapped in 2 Do not remove the clear pl...

Страница 12: ...bear in mind that the muscle contraction may cause involuntary limb movement which could cause injury to you or others Always ensure the limb is secured to prevent movement Turning On Off Receiver 1...

Страница 13: ...blinking on the Channel in use when it receives signal transmitting from the Remote Control 3 LED in Red Flashing battery power is low Receiver unit will turn on when charging the battery Orange light...

Страница 14: ...selected modality Note Make sure the identical receiver is turned on and properly placed For TENS programs When using any of the 8 programs for pain relief always start with the lowest intensity and g...

Страница 15: ...fortable tingling and pumping action action should appear to come in waves P8 for symptomatic relief and management of chronic intractable pain and relief of pain associated with arthritis Variable co...

Страница 16: ...d capillary supply It improves fatigue resistance during long duration moderate intensity exercise The exercise comprises an alternating sequence of work and rest phases lasting several seconds Increa...

Страница 17: ...The Well Life Wireless TENS EMS System is recommended for single person use at home Remote and Receivers The system device s may be wiped clean with a small amount of soapy water on a clean cloth Do n...

Страница 18: ...pt to disassemble or force open the built in battery Batteries have a finite lifetime for 12 months To maximize battery life please charge the system every 3 months when not in use If you find the bat...

Страница 19: ...harging Ensure the USB Cable is correctly into the Remote Receiver s and firmly plugged into the AC adaptor and into the electrical socket on the other side Receiver s does not pair with remote The re...

Страница 20: ...0 15A O P 5Vdc 1 2A Weight Dimension Remote Control 60 grams battery included 156 98x47x15 50 mm Wireless Unit 30 grams battery included 60x60x15 65 mm Operating Conditions 50 F 10 to 104 40 40 90 max...

Страница 21: ...quid or spray any liquids directly on power unit Stand by ii There is a label on the package explained as This symbol means used before represent as YYYY MM for year and month ELECTROMAGNETIC COMPATIB...

Страница 22: ...37 100 11 67 11 67 23 33 Declaration electromagnetic emissions and immunity for EQUIPMENT and SYSTEMS that are not LIFE SUPPORTING and are specified for use only in a shielded location The Wireless El...

Страница 23: ...ip in UT for 25 cycles 5 UT 95 dip in UT for 5 sec Mains power quality should be that of a typical commercial or hospital environment If the user of the EQUIPMENT or SYSTEM requires continued operatio...

Страница 24: ...off and on the user is encouraged to try to correct the interference by one or more of the following measures Reorient or relocate the receiving antenna Increase the separation between the equipment...

Страница 25: ...to be free from defects in workmanship and materials at the time of delivery The warranty does not apply to damage resulting from failure to follow the operating instructions accidents abuse alterati...

Отзывы: