Service Manual 610132 Rev. E
Pneumocheck Handle Printer/Charger 13
9
Depress & release test button: Ambient
temperature should be displayed.
20 degrees C = 68 degrees F
22 degrees C = 72 degrees F
24 degrees C = 75 degrees F
25 degrees C = 77 degrees F
27 degrees C = 81 degrees F
30 degrees C = 86 degrees F
If unit displays a reading outside ambient
temperature range, complete steps 10 thru 12.
Then, refer to troubleshooting guide. If unit
displays a reading of 390-400 degrees F,
check battery voltages. If battery voltage is
low, replace battery and repeat steps 4 thru
12.
10
Depress & release test button: Unit should
read FO = 55. This is a software coefficient.
11
Depress & release test button: Unit should
display a flashing ‘BAT OK’ message.
12
Depress & release test button: Unit should
display ‘DO DCV’ message.
13
Turn unit off. Depress test button and turn unit
on, release test button. Unit should read:
‘Test Mode’ >VER.3.5 >clear > select. The
latest software is VER 3.5: If unit reads VER
3.2, the software must be changed. The
PROM is part no 610102-535. It goes into
position U4 of the LCD digital PC board.
Section 2 - Service
Содержание PneumoCheck 61000
Страница 3: ...2 Pneumocheck Handle and Printer Charger Service Manual 610132 Rev E ...
Страница 39: ...38 Pneumocheck Handle and Printer Charger Service Manual 610132 Rev E ...
Страница 65: ...64 Pneumocheck Handle and Printer Charger Service Manual 610132 Rev E ...
Страница 69: ...68 Pneumocheck Handle and Printer Charger Service Manual 610132 Rev E ...
Страница 70: ......
Страница 71: ......
Страница 72: ......
Страница 73: ......
Страница 74: ......
Страница 75: ......
Страница 76: ......
Страница 77: ......
Страница 78: ......
Страница 79: ......
Страница 80: ......
Страница 81: ......
Страница 82: ......
Страница 83: ......
Страница 84: ......
Страница 85: ......
Страница 86: ......
Страница 87: ......
Страница 88: ......
Страница 89: ......
Страница 90: ......