46
Printing 226-Hz results
Welch Allyn MicroTymp3 portable tympanometric instrument
Print interpretive messages (Switch #2)
Print tympanogram only or tympanogram and data (Switch #3)
Manufacturing switch (Switch #4)
No messages
Depress the ON portion of the switch to select this option.
This causes messages which interpret the tympanogram to
not be included on the printout.
Refer to the Printout formats section on
page 42
for more
information on these messages.
Messages
Depress the OFF portion of the switch to select this option.
This causes the messages which interpret the
tympanograms to be included on the printout.
OFF
1
2
3
4
1
2
3
4
OFF
Tympanogram only
Depress the ON portion of the switch to select this option.
Only the tympanogram and the GR (Width) numeric value will
print.
Tympanogram and data
Depress the OFF portion of the switch to select this option.
Both the tympanogram and its corresponding numeric data
will print.
1
2
3
4
OFF
1
2
3
4
OFF
This switch is used during manufacturing only. Leave this
switch in the OFF position. The Printer/Charger will not
operate normally if this switch is on.
1
2
3
4
OFF
Содержание MicroTymp 3 Handle
Страница 1: ...MicroTymp 3 portable tympanometric instrument Directions for use...
Страница 6: ...vi Contents Welch Allyn MicroTymp3 portable tympanometric instrument...
Страница 16: ...16 Introduction Welch Allyn MicroTymp3 portable tympanometric instrument...
Страница 22: ...22 Completing a test Welch Allyn MicroTymp3 portable tympanometric instrument...
Страница 30: ...30 Understanding the LCD messages Welch Allyn MicroTymp3 portable tympanometric instrument...
Страница 48: ...48 Printing 226 Hz results Welch Allyn MicroTymp3 portable tympanometric instrument...
Страница 72: ...72 Specifications Welch Allyn MicroTymp3 portable tympanometric instrument...