24
Understanding the LCD messages
Welch Allyn MicroTymp3 portable tympanometric instrument
Open
The test has not begun since a valid ear cavity has not been detected.
Possible causes:
•
Instrument is not in an ear
•
Probe tip is not completely sealed in an ear
•
Perforated tympanic membrane
•
Ear with patent tympanostomy tubes
•
Extremely large canal (> 2.5 cc)
Possible solutions:
•
Reposition the probe tip
•
Perform otoscopy to check for perforated membrane or patent tympanostomy tubes
Test
The TEST message indicates that the test has started. Test data will begin to appear
immediately.
-400
-200
0
200
2
0.5
0
1.0
1
GR
A
GR
C
PRESSURE - daPa
Ya - mmho
+200 Vea - cc
-400
-200
0
200
2
0.5
0
1.0
1
GR
A
GR
C
PRESSURE - daPa
Ya - mmho
+200 Vea - cc
-400
-200
0
200
2
0.5
0
1.0
1
GR
A
GR
C
PRESSURE - daPa
Ya - mmho
+200 Vea - cc
-400
-200
0
200
2
0.5
0
1.0
1
GR
A
GR
C
PRESSURE - daPa
Ya - mmho
+200 Vea - cc
Содержание MicroTymp 3 Handle
Страница 1: ...MicroTymp 3 portable tympanometric instrument Directions for use...
Страница 6: ...vi Contents Welch Allyn MicroTymp3 portable tympanometric instrument...
Страница 16: ...16 Introduction Welch Allyn MicroTymp3 portable tympanometric instrument...
Страница 22: ...22 Completing a test Welch Allyn MicroTymp3 portable tympanometric instrument...
Страница 30: ...30 Understanding the LCD messages Welch Allyn MicroTymp3 portable tympanometric instrument...
Страница 48: ...48 Printing 226 Hz results Welch Allyn MicroTymp3 portable tympanometric instrument...
Страница 72: ...72 Specifications Welch Allyn MicroTymp3 portable tympanometric instrument...