Section 4.2 Disassembly & Reassembly
Re-assembly and Repair of the Handle
P
3. PUT PUMP INTO HOUSING.
WHEN PUTTING THE PUMP INTO HOUSING, MAKE SURE ALL FOUR PUMP
PADS ARE POSITIONED PROPERLY. THE FIRST WILL BE ON THE TOP CAP
ASSEMBLY. (SOMETIMES IT WILL STICK TO THE UNDERSIDE OF THE
FRONT HOUSING PATIENT SIDE). LOOK FOR IT THERE IF IT NOT ON THE
TOP OF THE TOP CAP. TWO SMALLER PADS WILL BE LOCATED AT THE
FRONT OF THE PUMP ON BOTH SIDES OF THE MOTOR. THE FOURTH PAD IS
ON THE BOTTOM OF THE INSIDE BACK HOUSING. WHEN PUTTING THE
PUMP BACK INTO THE HOUSING, MAKE SURE THAT THIS FOURTH PAD IS
BETWEEN THE FLEX CIRCUIT AND THE BACK OF THE PUMP. BE SURE
THAT THE OTHER PADS STAY IN PLACE WHEN PUTTING THE PUMP BACK
INTO THE HOUSING.
Q
1. ATTACH FRONT HOUSING (PATIENT SIDE HOUSING).
WRAP THE HOSE AROUND THE PAD ON THE PRESSURE TRANSDUCER
UNTIL THERE IS NO SLACK IN THE LINE WHEN THE HOUSING IS ALMOST
COMPLETELY CLOSED. PUT THE HOUSING IN PLACE MAKING SURE NOT
TO PINCH THE HOSE. TIGHTEN THE TWO SCREWS. CHECK THE HOUS-
INGS TO MAKE SURE THEY ARE MATED UNIFORMLY AROUND ALL EDGES.
WIPE THE MOUNTING AREA FOR THE LABEL WITH A SWAB DAMPENED
WITH ISOPROPYL ALCOHOL. WAIT APPROXIMATELY ONE MINUTE FOR
THE ALCOHOL TO EVAPORATE. ATTACH A NEW WELCH ALLYN LABEL.
R
1. SECURE HOUSINGS TOGETHER WITH LOCTITE BLACK MAX.
CAUTION WHEN USING ADHESIVES:
ALWAYS WEAR SAFETY GLASSES AND PROVIDE ADEQUATE VENTILA-
TION WHEN USING ADHESIVES OR ACCELERATORS. READ THE MSDS
SHEET BEFORE USING THESE MATERIALS. COMPLY WITH ALL SAFETY
PROCEDURES WHILE USING THESE MATERIALS. BE CAREFUL NOT TO
GET BLACK MAX BETWEEN THE FLEX CIRCUIT AND THE TAB. ALSO DO
NOT GET IT INTO THE GROOVE THAT THE BATTERY COVER SITS IN OR
ON THE EXTERIOR SURFACES OF THE MICROTYMP.
USE A SMALL APPLICATOR TIP ON THE END OF THE BLACK MAX BOTTLE.
APPLY A SMALL DOT ON THE SURFACE OF THE BLACK TAB THAT FACES
THE INSIDE OF THE HOUSING. DO THIS TO THE LEFT SIDE AND RIGHT
SIDE OF THE MICROTYMP. APPLY A DROP OF ACCELERATOR ON THE
BLACK MAX WHILE PRESSING THE HOUSINGS TOGETHER. HOLD FOR
ONE MINUTE TO ALLOW THE BLACK MAX TO FULLY CURE.
pg 51
Содержание Micro Tymp 1
Страница 10: ......
Страница 11: ......
Страница 12: ......
Страница 22: ...Section 2 4 Service MT 1 CALIBRATION PROCEDURE CALIBRATION NOTES date____ ____ ____ pg 19...
Страница 47: ......
Страница 51: ......
Страница 52: ......
Страница 54: ......
Страница 56: ......
Страница 67: ......
Страница 68: ......
Страница 69: ......
Страница 70: ......
Страница 71: ......
Страница 72: ......
Страница 73: ......
Страница 74: ......
Страница 75: ......
Страница 76: ......
Страница 77: ......
Страница 78: ......
Страница 79: ......
Страница 80: ......
Страница 81: ......
Страница 82: ......