Attach the probe tube
A clear probe tube must be attached to the probe head before an eartip is applied.
1. Insert a new probe tube into the probe head until it is fully seated.
A properly inserted probe tube snaps securely into place when it is fully seated in the
probe head.
2. To remove a probe tube, grasp the tube and pull gently away from the probe head
with a slight twist.
It is not necessary to replace the probe tube with each eartip; the tube is reusable as
long as the probe tube is clear.
WARNING
If the probe tube becomes dirty or clogged, it must be
replaced. A dirty probe tube may produce wrong results.
Eartips
The screener comes with a box of disposable, single-use eartips that fit a variety of ear
canal sizes. The determination of the appropriate eartip size should be made by persons
with proper training and experience. The probe must have a probe tube applied and an
eartip attached before inserting it into an ear canal. The eartip must seal the ear canal.
The best test results are obtained when the eartip is inserted deeply into the ear canal
instead of flush with the ear canal opening. Caution must be taken, however, to ensure
that the eartip does not extend too deeply into the ear canal. Use only the eartips
approved for use with the instrument. The eartips are disposable and must be replaced
after each patient.
Attach and remove the eartips
WARNING
A clear probe tube must be applied to the probe head before
an eartip is applied.
WARNING
The eartips are disposable and must be replaced after each
patient. Failure to do so may cause infection.
WARNING
The screener probe must have a probe tube and an eartip
attached before inserting the screener probe into the ear canal. Failure to
do so may cause infection.
WARNING
A person with proper training and experience must determine
the appropriate eartip size. Failure to do so may cause infection.
14 Setup
Welch Allyn
®
OAE Hearing Screener
Содержание 35900-NP-TE
Страница 1: ...Welch Allyn OAE Hearing Screener Directions for use ...
Страница 10: ...6 About warnings and cautions Welch Allyn OAE Hearing Screener ...
Страница 14: ...10 Introduction Welch Allyn OAE Hearing Screener ...
Страница 22: ...18 Setup Welch Allyn OAE Hearing Screener ...
Страница 40: ...36 Manage results Welch Allyn OAE Hearing Screener ...
Страница 46: ...42 Maintenance Welch Allyn OAE Hearing Screener ...
Страница 56: ...52 Standards and compliance Welch Allyn OAE Hearing Screener ...
Страница 58: ...54 Warranty Welch Allyn OAE Hearing Screener ...
Страница 62: ...Appendix B Flowchart Measurement 58 Appendices Welch Allyn OAE Hearing Screener ...
Страница 71: ......