Troubleshooting the Sprint 
 

What do you need in a kit to troubleshoot the Sprint? 

 

Batteries (2xAA) 

transmitting cable 

headset cable 

Signal check  (part no. Z22502) 

lapel microphone  

Monitor earphones for the sprint (part no. Z43139) 

CL clinic contact numbers 

Guide to codes on LCD 

 

Check

The low battery symbol is not showing. 
The correct coil is properly in place.  (Coil should have ‘S’ or ‘L’ 

embossed on to it.) 
Both the headset cable and transmitting cable are connected 

properly. 
The cables are not broken or cracked. 
 
When the processor is working correctly 5 – 8 
segments will show on the segment meter in 
response to speech.  The red indicator light will also 

flash.   
 
 

If the indicator light and segment meter do not respond to 
sound: 

 

Change all batteries ensuring that they are in the correct way. 

 

Ensure that battery contacts are clean. Rub the terminals with 
an eraser and clean with alcohol wipes if possible. 