this may additionally heat up the respir-
atory air and inside parts. This might
cause condensation to form in the hu-
midifier and to be deposited in the hose
system.
•
Do not cover the unit with blankets etc.
•
Do not transport SOMNO
vent
S with
the humidifier installed. Tilting might
cause water remaining in the humidifier
to run into the SOMNO
vent
S and dam-
age it.
•
Do not place a mobile phone when
switched on less than 1 m from
SOMNO
vent
S as this might result in
malfunctions.
•
Only have checks and repair work car-
ried out by Weinmann, the manufactur-
er, or specialist staff.
•
If third-party items are used, functional
failures may occur and fitness for use
may be restricted. Biocompatibility re-
quirements may also not be met. Please
note that in such cases, any claim under
warranty and liability will be voided if
neither the accessories nor genuine re-
placement parts recommended in the
instructions for use are used.
•
Third-party masks may only be used af-
ter being approved by Weinmann, the
manufacturer.
•
Please observe the section "6. Hygienic
preparation" on page 48 to avoid infec-
tion or bacterial contamination.
•
If a pneumotachograph with a high flow
resistance is used to determine the flow
when initiating or controlling therapy,
the trigger function of SOMNO
vent
S
may be impaired. In the case of queries
please contact Weinmann, the
manufacturer.