Problem

Possible Cause

Solution

Leg pain after use 

of device

You may have set the 

intensity too high

Adjust and set the intensity at a lower 

level (where you can feel the mild elec

trical stimulation & it is comfortable)

The device turns on, 

but does not generate 

electric impulse

The device is not being 

used correctly

Check whether your hands can feel the 

stimulation. To test this, place both your 

bare hands on both the footpads and 

increase the intensity. If you feel the 

stimulation then the device is working 

fine but if you do not feel it even at the 

highest intensity then please contact 

your authorised dealer.

Not placing both bare 

feet on the foot pads 

at the same time

Ensure that both your feet are bare and 

they are placed properly and 

simultaneously on the foot pads. Then 

increase the intensity until you feel the 

stimulation.

The intensity level may 

be too low

Keep increasing the intensity and take it 

to the maximum level until you feel the 

stimulation.

Do not compare yourself with other 

people who feel the stimulation at lower 

level of intensity because everyone is 

different.

Electrodes patches 

cable broken

Replace the cable

Electrodes patches 

cable not connected 

properly

Connect the cable properly

You are using this 

device too frequently

Keep sufficient gap between two 

consecutive treatments to allow the 

muscles to recover. Reduce the duration 

until your muscles have acclimatized to 

the stimulation

13. Troubleshooting

Problem

Possible Cause

Solution

The device turns on, 

but does not generate 

electric impulse

Your feet may be dry

Moisturize the soles of your feet to 

improve conductivity and stimulation 

and try the procedure again. You may 

also have to increase the intensity level.

The device does not 

turn on/ No indication 

of light

Batteries not recharged

Recharge the batteries

The device turns on 

and then goes off 

again

Batteries not recharged

Recharge the batteries

Blinking red light 

when trying to turn 

on the device.

Low batteries.

Recharge battery.

The device does not 

turn on even after 

new batteries have 

been inserted

contact the authorized dealer

The device does 

not vibrate

The device is not 

designed to vibrate

You may be dehydrated Drink plenty of water before and after 

using the device. The device uses your 

body to create the electrical circuit. 

Water is an excellent conductor of 

electricity and if your body is less 

hydrated (below 60%) then the 

stimulation may be less, therefore 

it is important to always remain well 

hydrated.

Содержание CIRCULATOR

Страница 1: ...16 Plot Number Sector 2 Akshar Industrial Park Opp Zydus Cadila Pharma Sarkhej Bavla National Highway Moriaya Ahmedabad 382213 Gujarat India Manufactured For COMFORT CLICK LIMITED Number 7 Doctor Cro...

Страница 2: ...ation 9 Care Cleaning and maintenance 10 Purchasing consumables 11 Expected service life 12 Disposal of product WEEE 13 Troubleshooting 14 EMC and Electrical standard requirement 8 9 10 11 11 11 12 13...

Страница 3: ...irculation To relieve neuropathy pain osteoarthritic pain acute and chronic body pain varicose vein To provide paralytic nervous activation For strengthening body muscle and relieve body muscle pain F...

Страница 4: ...ef by exciting sensory nerves and thereby stimulating either the pain gate mechanism and or the opioid system The type of stimulation delivered by the TENS unit aims to excite stimulate the sensory ne...

Страница 5: ...ulation over neck there is a chance that using the electrode patches on neck may cause the closure of airway dif culty in breathing Stimulation over the neck could adversely impact hearing or blood pr...

Страница 6: ...cables periodically damage Do not spill liquid or water on the device Do not use this device with Electrode Patches Cable Electrode Patches other than those provided by Camex Wellness Limited The per...

Страница 7: ...g with any other types of DC USB Cable may damage the device Usage of any other attachment that is not with your charger may result in electric shock or injury to people Shortening the terminal of the...

Страница 8: ...tex free Electrode patches Clean the foot pads with a wet clothe and dry cleaning brush The foot pads are latex free Circulator Nerve and muscle stimulator Do not clean the device while charging 10 Pu...

Страница 9: ...se everyone is different Electrodes patches cable broken Replace the cable Electrodes patches cable not connected properly Connect the cable properly You are using this device too frequently Keep suf...

Страница 10: ...ode patches have been over used or are dry Replace electrode patches Sensation is too strong Setting is too high Press the button to decrease the setting Sensation is too weak Sensation is too low Pre...

Страница 11: ...gress of water with harmful effects IPX8 Protection against continuous immersion in water up to 1 meter or as speci ed by user ON power General Warning Sign Country of manufacture To identify the coun...

Страница 12: ...e product within the applicable warranty period If there is not a Distributor you must contact Camex Wellness Limited directly 20 Device foot pads and foot intensity The device should be charged for a...

Страница 13: ...he device Step 2 Adjust the intensity through foot intensity level buttons Continue to increase the foot intensity at your comfort level Step 3 As the device is set on Auto Timer for 30 mins post 30 m...

Страница 14: ...es Peel off the protective membrane from the electrode patches and carefully place on the clean dry uninjured skin Step 2 Turn on the device through the power button of device Step 3 Adjust the intens...

Страница 15: ...10 seconds during the therapy in the PROG TIME display segment Information The poly fuse has been used in the devices and that is resettable We uses on board SMD 1812L150 LF150 Fuse Note Camex wellne...

Страница 16: ...on Date 20 06 2022 Document No CWL URM 04A Revision No 00 Revision Date 20 06 2022 29 30 EMS Voltage Waveform Foot Body 12 85v 8 58v Section Voltage Range Foot Body 20 120v 12 84v Section Voltage Rang...

Страница 17: ...Document No CWL URM 04A Revision No 00 Revision Date 20 06 2022 Document No CWL URM 04A Revision No 00 Revision Date 20 06 2022 31 32...

Отзывы: