![Weidmüller PTX800A Скачать руководство пользователя страница 45](http://html1.mh-extra.com/html/weidmuller/ptx800a/ptx800a_operating-instructions-manual_952251045.webp)
45
ChangetheInputtypeinthesoftware:
1. Power up the instrument and start the setup routine.
. While the software version number is flashing, remove the security
link and press
PGM
.
. The display will show a number as shown in the table below.
Number
0
1
2
3
4
5
InputType
mA
V
mA
V
mA
V
Outputtype
mA
mA
V
V
None
None
4. Change the number to suit your input and output types.
5. Replace the security link and press
PGM
.
Changetheoutputrangesettings:
6. Scroll through the setup and change the input type and range
settings - you will be required to calibrate the Inputs.
Calibratetheinput:
. Calibrate the inputs to the new output range and type.
R3
Analogue
Output
Board
R1
R2
D
ispla
y B
oar
d