Procedure
Whenthe
displayshows
Action/Description
Put the instrument in setup mode and scroll through the main menu
CALIn_n
Press
ì
or
Å
CAlIn_y
Press
PGM
to select Input calibration
InP_LO_=
Connect the multimeter to measure the Input level, then press
PGM
4.00
Set the Input signal to the appropriate value, e.g., 4.00 mA
When you are happy that the Input is correct, press
PGM
Inp_HI_=
After approx 5s the display introduces the next calibration point. Press
PGM
20.00
Set the Input signal to the appropriate value, e.g., 20.00 mA
When you are happy that the Input is correct, press
PGM
dFactor=
Continue with the setup sequence.
Note: Do not remove the power while the save message is on display.
PTX800 analogue output board showing output type jumper locations.
Link both pairs
of posts for
voltage outputs