connecting the sensors
3.2
HPM4030 with Webtec can sensors
CAN cable
SR-CBL-02-MF-CAN
CAN terminating
resistor
SR-CBL-000-R-CAN
CAN Y-junction
SR-CBL-0.3-Y-
CAN
The SORT menu option can be used to change the order of display (see page 35).
The sorting order can be reset using the Reset SORT option (see page 36). The sensors are then sorted in the order that they are
connected.
The CAN sensors are assigned to channels
in the order that they are connected. This
sorting order is maintained even while the
instrument is turned off.
Make sure that the CAN bus is terminated on
the last sensor using a SR-CBL-000-R-CAN
terminating resistor.