Broadband Analyzer/Generator
ANT-20/ANT-20E
S-2
Specifications “ATM Layer”
1.4
General Functions
1.4.1
Error Insertion (Anomalies)
The following anomalies can be inserted in addition to the errors inserted in the physical layer.
These anomalies are not dependent on the active measurement type, they are referred to the
entire cell stream including fill cells.
1.4.2
Alarm Generation (Defects)
The following defects can be generated in addition to the physical layer alarms. These defects
can be generated independently of the active measurement type. They are referred to the entire
cell stream.
Error type,
anomaly
Single
Rate
1
Sensor
threshold
M in N
HEC uncor.
2
yes
1E-2 to 1E-6
M=1 to 31
N = M+1 to M + 31
HEC cor.
3
yes
1E-2 to 1E-6
M =1 to 31
N = M +1 to M + 31
1 Mantissa = 1 (fixed), exponent = -2 to -6 (integer values)
2 Uncorrectable header error
3 Correctable header error
Table S-1
Settable anomalies
Defect
Test sensor functions
Single
On/Off
LCD
1
yes
yes
1 LCD (Loss of Cell Delineation) is generated if uncorrectable header errors are found
in seven or more consecutive cells.
Table S-2
Settable defects