background image

16

6.2 Cuvette Replacement

HF scientific recommends that the cuvette be replaced annually or if it 

appears badly soiled or discolored at any time. It is suggested to place 

a covering over a floor grate if the instrument is installed over a grate.
To replace the cuvette:
Step 1:   Press SERVICE to stop the flow of sample water and drain 

the cuvette. 

Step 2:   Ensure the source water is turned off.

Step 3:   Turn counterclockwise, but do not completely remove, the 

knurled fitting on top of the optics system to loosen the 

cuvette. Occasionally it is necessary to very gently rotate the 

cuvette on the O-rings to facilitate removal. Once the cuvette 

rotates easily on the O-rings carefully remove the cuvette. 

There is a spring installed in the optical block to facilitate 

removal of the cuvette.

Step 5:   Reverse the procedure to install a new or cleaned cuvette. 

Step 6:   Retain the removed cuvette for future use if it can be cleaned.
Refer to Figure 5 - Wet Chemistry Area of Enclosure
Install the new cuvette by pushing it firmly but gently to seat it on the 

O-rings, carefully turn the knurled top clockwise until the cuvette is held 

level and securely against both the top and bottom O-rings, do not 

overtighten. Return to normal operation.

6.3 T-Strainer Cleaning

The T-strainer is integral to the instrument and must be checked when 

changing the reagent. When necessary the T-strainer screen and bowl 

must be cleaned. The T-strainer is clamped to the door and does not 

require removal to be cleaned. You may want to place a container 

under the instrument to catch the water and debris during cleaning.
Step 1:  Press SERVICE to stop the flow of sample water and drain 

the cuvette. 

Step 2:  Ensure the source water is turned off.

Step 3:   Disconnect the top (bowl) of the T-strainer by turning 

counterclockwise. 

Step 4:   Remove and clean the screen and the bowl and then 

reassemble.

Step 5:   Be sure to tighten the bowl of the strainer. 

Step 6:   Turn source water back on.

Step 7:   Return to normal operation. 

Step 8:   Check for any leaks.
The T-strainer screen (Cat. No. 28625) may require replacement after a 

period of time or as needed.

6.4 Adjusting the Pressure Regulator

The pressure regulator is factory set and it should not be necessary to 

adjust it in the field.

6.5 Installing New Reagent in the SSR-Ex

Use caution when changing the reagents, as they are corrosive. These 

reagents may stain clothing. After changing the reagents, operators 

should wash their hands.
Step 1:   Press SERVICE to stop the flow of sample water and drain 

the cuvette. 

Step 2:   Ensure the source water is turned off.

Step 3:   Remove both reagent bottles by unscrewing cap assembly. 

Step 4:  The buffer is installed on the right and the indicator is installed 

on the left.

Step 5:   Once bottles are secure in the bracket, wrap the Velcro 

around both bottles to ensure they do not move.

Step 6:   Screw cap on the bottle and ensure a tight fit. Ensure tubing 

is fully submerged at the bottom of the bottle. 

Step 7:   Follow the PRIME procedure under the SERVICE menu. 

This will draw enough of each reagent to completely prime 

the tubes and replace any old solution. The system will 

automatically return to normal operation after it completes the 

prime.

6.6  Replacing the Peristaltic Sample  

Pump Head

The pump head assembly can be replaced. Recommended replacement 

for the pump head assembly is annually. It is replaced as a complete 

assembly, not just the pump tube. The pump is located in the upper 

right of the wet enclosure.
Step 1:   Press SERVICE to stop the flow of sample water and drain 

the cuvette. 

Step 2:   Ensure the source water is turned off.

Step 3:   Squeeze the right side of the pump housing, this will release 

the pump head from the bracket and a new pump head can 

be installed.

Step 4:   Disconnect the inlet and outlet sample tubes from the pump 

once it has been removed from the pump base taking care to 

note which tube connects to which side of the pump head.

Step 5:   Connect the inlet and outlet sample tubing to the new pump. 

Step 6:   Carefully replace the pump head in the enclosure.

6.7 Instrument Storage

If the SSR-Ex is relocated or will be inactive for long periods of time 

(several months). 
Step 1:   Press SERVICE to stop the flow of sample water and drain 

the cuvette.

Step 2:   Ensure the source water is turned off.

Step 3:   Remove the reagents.

Step 4:   Flush the reagent sample lines with distilled water. Flush the 

reagent system. 

Step 5:   Remove power by disconnecting the mains power.

Step 6:   It is usually a good idea to disconnect or shut off the source 

water for the duration of storage.

To order any accessory or replacement part, please contact the HF 

scientific Customer Service Department. If for any reason technical 

assistance is needed regarding this instrument, please do not hesitate 

to contact the HF scientific Technical Services Department.

Содержание HF scientific SSR-Ex 28037

Страница 1: ...Model SSR Ex 28037 IOM HF SSR Ex 28037 WARNING Read this Manual BEFORE using this equipment Failure to read and follow all safety and use information can result in death serious personal injury property damage or damage to the equipment Keep this Manual for future reference IOM HF SSR Ex 28037 Rev 5 ...

Страница 2: ...verage Menu 13 5 2 1 8 Backlight Menu 13 5 2 1 9 Cycle Time Menu 14 5 2 1 10 Contact Period Menu 14 5 2 1 11 Standby Pulse Menu 14 5 2 1 12 Restore Defaults Menu 14 5 2 1 13 Restore Factory Cal Menu 14 5 2 1 14 About Device Menu 14 5 3 System Start Up 15 5 3 1 Flush Water Sample Line 15 5 3 2 Prepare Reagents 15 5 3 3 Remote Start 15 5 4 Typical Cycle 15 5 5 Observe Operation 15 6 0 Maintenance 15...

Страница 3: ...d the Manual This pictorial alerts you to electricity electrocution and shock hazards 1 0 Safety 2 0 Overview This instrument has been specifically designed for TRO measurement and to operate in hazardous atmospheres meeting the IECEx and ATEX ratings as stated in the specifications The instrument uses an IECEx approved purge pressurization controller called the Air Sentinel II Please refer to the...

Страница 4: ...ve FKM Polypropylene 316 Stainless Steel Acetal EPDM Silicone Modified PPO Thermoplastic Elastomer Sample Temperature Range 0 C 55 C 32 F 131 F Power Supply 100 to 230VAC 47 63 Hz 240VA Insulation Rating Double Insulated Pollution Degree 2 Overvoltage Category II Enclosure Powder coated steel with windows IP55 rating Environmental Conditions Altitude up to 2000 meters Up to 95 RH non condensing IE...

Страница 5: ...nside Front of Electrical Panel Enclosure Warning and informational Front of Unit Informational Front of Unit 2 3 Instrument Labels The following labels should be applied to the outside of the enclosure 3 0 Components Listing 3 1 Electronics Compartment WETTED COMPONENTS COMPARTMENT ACCESSIBLE ANY TIME CONTAINS REAGENT BOTTLES OPTICAL MEASUREMENT COMPONENTS INLET WATER SAMPLE PUMP PERISTALTIC CONN...

Страница 6: ...e for use with 6mm tubing if using the supplied fittings for installation ensure that a proper thread sealant is used Element Replacement Indicator The red ring will pop up to show when the element needs to be replaced Figure 7 Air Dryer Filter 4 0 Installation 4 1 Mounting and Site Selection The instrument is designed for wall mounting Choose a location that is easily accessible for operation and...

Страница 7: ...valve should be located close to the instrument to allow for periodic maintenance EMERGENCY DRAIN NO CONNECTION NEEDED SAMPLE WATER 6mm connection MAIN DRAIN 12 mm connection PURGE AIR AIR ONLY 6mm connection NOTE IMAGE DEPICTS BOTTOM OF THE ENCLOSURE DO NOT MIX UP THE AIR and WATER CONNECTIONS INSTALLING THE WATER INTO THE AIR CONNECTION WILL VOID THE WARRANTY Figure 10 Utility Connections All we...

Страница 8: ...n every bus may require a 120 ohm termination resistor to eliminate the possibilities of signal reflection on the line Do not run RS 485 cables in the same conduit as power Ensure each instrument is not powered when connecting the RS 485 line To prevent damage to the instrument ensure that power is disconnected prior to making connections The Modbus connection is located on the lower right portion...

Страница 9: ...he reading if True False 00007 Prime mode Start a prime if True False 4 5 2 Input Status These single bit values are readable from the master The data will be returned with the lowest addressed input status in the least significant bit LSB of the data Unused bits in the data will be set to 0 4 5 2 1 Valid Command s Code Name Broadcast 0x02 Read Input Status No 4 5 2 2 Format 16 bit word format MSB...

Страница 10: ... mantissa 4 5 4 3 Valid Addresses 30001 30XXX Address Type Register Function 30001 30002 Float Sensor read ing The meter reading 30003 30004 Float ABS The absorbance 30005 Int Main version Main board software version 30006 Int Display version Display board software version 30007 Int NOT USED 30008 Int Model number 28037 30009 Int Model suffix number Options model dependent 0 30010 Int Reading stat...

Страница 11: ...R Ex The SSR Ex is configured using the buttons located on the outside of the electrical enclosure The Mainscreen is shown below in Figure 12 Main Screen Using the UP and DOWN buttons navigate to the desired function shown on the screen Once you have highlighted the desired function press SELECT to display the next menu The Service and Config menus permit access to the processes listed in the char...

Страница 12: ...resentative sample to the SSR Ex Every time SERVICE mode is exited all faults are cleared If the original fault or a new fault occurs it will be posted to the display Faults will not stay cleared until the fault condition is no longer detected Select using UP DOWN buttons HOME No Change PRIME CALIBRATE WATER PRIME SEE BELOW CONFIG ANALOG OUTPUT see 4 2 1 3 CONFIG MODBUS see 4 2 1 4 ALARMS see 4 2 ...

Страница 13: ...S ANALOG OUTPUT MODBUS ALARMS 5 2 1 4 Modbus Menu 5 2 1 5 Alarms Menu The factory default setting is 9600 8 bit no parity 1 stop bit RTU All parameters are adjustable Make sure the parameters on the Host and the SSR Ex are configured identically Select using UP DOWN buttons These settings allow the user to configure the SSR Ex MODBUS Settings include the MODE BAUD RATE ADDRESS PARITY and BITS Sele...

Страница 14: ...n OFF Default setting is ON and default time is 24 Hr This option will restore the factory default settings Screen displays RESET DEFAULTS Pressing the SELECT button will restore factory default settings Before performing a Restore Defaults it is advised to make a listing of any user changed menu options prior to the reset 5 2 1 12 Restore Defaults Menu 5 2 1 11 Standby Pulse Menu RESTORE DEFAULTS...

Страница 15: ...e enclosure needs to be accessed for service ensure all power has been removed from a safe location prior to accessing the electrical or powered devices in the electrical enclosure All components in the lower wetted parts section of the enclosure can be accessed for service or routine Figure 13 Pressure regulator and T strainer 6 1 Calibration Calibration kit Cat No 28144S includes 125 ml of zero ...

Страница 16: ... when changing the reagents as they are corrosive These reagents may stain clothing After changing the reagents operators should wash their hands Step 1 Press SERVICE to stop the flow of sample water and drain the cuvette Step 2 Ensure the source water is turned off Step 3 Remove both reagent bottles by unscrewing cap assembly Step 4 The buffer is installed on the right and the indicator is instal...

Страница 17: ...olatile error Try factory reset Main Board Lost Internal communication loss Check cable connections try factory reset Low Setpoint Alarm Reading below set point Check source water oxidant level High Setpoint Alarm Reading above set point Check source water oxidant level 10 0 Spare Parts Listing Part Kit Name Description Frequency of part or kit change Calibration Kit 28144S Description Includes 12...

Страница 18: ..._ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ ...

Страница 19: ..._ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ ...

Страница 20: ... replacing other property which is damaged if this product does not work properly other costs resulting from labor charges delays vandalism negligence fouling caused by foreign material damage from adverse water conditions chemical or any other circumstances over which the Company has no control In addition the Company shall not be responsible for any costs incidental to the Company s warranty res...

Отзывы: