4.5.5 Exception Responses Implemented
5.0 Configuration - General
Process
5.1 Instrument Commissioning and
Start-up
After all water, air, and electrical connections have been made, the
SSR-Ex can be powered on by turning on the power to the Air Sentinel
II. Once power is supplied to the Air Sentinel II, it will run through an
automatic three-minute purge cycle and then supply power to the
SSR-Ex. If at any time there is a loss of pressurization in the electrical
enclosure, the Air Sentinel II will remove power to the SSR-Ex.
The condition of the Air Sentinel II can be determined based on the
color of the LED lamps on its body as explained by this chart:
Lamp Color Matrix Meaning
Power to
Attached
Device
Alarm Relay
Response
Both Lamps
GREEN
Safe
Operation
ON
Normally Open
Contacts
OPEN
TIMER Lamp
RED
Instrument
is Purging
OFF
Normally Open
Contacts
CLOSE
PRESSURE Lamp
RED
Unsafe
Operation
OFF
Normally Open
Contacts
CLOSE
NO Lamps
No power
out applied
No power
applied;
everything
off
The SSR-Ex is ready for operation when the Air Sentinel II displays two
green LED lamps.
5.2 Configuring the SSR-Ex
The SSR-Ex is configured using the buttons located on the outside of
the electrical enclosure. The Mainscreen is shown below in Figure 12 -
Main Screen.
Using the UP and DOWN buttons, navigate to the desired function
shown on the screen. Once you have highlighted the desired function
press
SELECT to display the next menu.
The Service and Config menus permit access to the processes listed in
the charts below. Use the EXIT button to return to the previous screen.
Figure 12 - Main Screen
Service Menu Options
Config Menu Options
Title
Function/Description
PRIME
Prime DPD and Buffer Reagents.
CALIBRATE
Initiate Calibration Procedure. Follow
prompts to complete.
WATER PRIME
Actuate the sample pump to flush the
water lines.
Title
Function/Description
ANALOG OUTPUT
Configure 4 - 20 mA Output
MODBUS
Configure MODBUS
ALARMS
Configure Alarms
UNITS
Select PPM or mg/l
AVERAGE
Select Signal Averaging 1 - 5 cycles
BACKLIGHT
Adjust Display Backlighting
CYCLE TIME
Reading Update Rate
CONTACT PERIOD
Permits Chlorine Contact time before
measuring
STANDBY PULSE
Pulses reagent to reduce clogging
while in standby
RESTORE DEFAULTS
Restores factory default settings
RESTORE FACTORY CAL.
Restore factory calibrations
ABOUT DEVICE
Displays the DISPLAY and MAIN
software version numbers