background image

15

5.2.1.13 Restore Factory Cal Menu

5.3 System Start-up

Once the SSR-Ex is powered on, air has been purged, and all 

configurations have been completed, it is ready for use.

5.3.1 Flush Water Sample Line

Before sampling any sample water, the water line from the source 

water to the enclosure must be flushed to ensure that the sample line 

has a representative sample and is not impacted by any debris or other 

foulants. Flush the sample line by using the WATER PRIME feature 

found under the SERVICE menu. When this feature is selected, the 

connect- ed water sample line will automatically flush water to drain.
Reference section 4.2.1.1 main menu.

5.3.2 Prepare Reagents

Prepare and install the reagents according to the instructions provided 

on the reagent bottles. In the SERVICE menu, select PRIME to prime 

the reagents. You can access the bottom of the enclosure to service 

these items, if needed.

5.3.3 Remote Start

Once the line is flushed and the reagents primed, initiate a cycle 

remotely through RMT or MODBUS.

The SSR-Ex will only operate using a remote start signal.

5.4 Typical Cycle

During normal operation, the SSR-Ex will run through a timed cycle. A 

simplified cycle will consist of the following sequences:
Flushing – sample flow Purging – PURGE valve opens

Zeroing – no flow with cuvette full

Adding Reagents – one pulse of the reagent pumps

Sample Reaction – reaction of reagents with oxidant sample Reading 

resulting sample – no flow with cuvette full

Purging – PURGE valve opens to remove the reacted sample and water
The cycle above is simplified and does not describe all the actions 

and testing that occurs. The CPU continuously diagnoses the entire 

system for correct operation and sample water flow. If an error occurs, 

a message is posted on the LCD screen and is indicated on a Modbus 

address. Alarm contact closure can be user configured to provide an 

alarm in the event of an error message.

5.5 Observe Operation

Observe the process in the optical assembly, you should see a vortex 

of water form as the SSR-Ex flushes water through the cuvette, then 

the unit will stop water flow and “zero” on the source water, flush again 

and then actuate the reagent pumps and take

6.0 Maintenance 

Under normal operating conditions, it is recommended that the 

main enclosure door is closed. If the electrical or power devices in 

the purged (upper) section of the enclosure needs to be accessed 

for service, ensure all power has been removed from a safe location 

before accessing the electrical or powered devices in the electrical 

enclosure.

The SSR-Ex instrument is sectioned such that all the electrical or powered 

devices are in the upper, purged section of the enclosure, while the lower 

(wetted parts) section contains no power at all and can be operated and 

maintained in a safe condition even with the main enclosure door open. In 

the interest of safety, if the electrical or power devices in the purged (upper) 

section of the enclosure needs to be accessed for service, ensure all power 

has been removed from a safe location prior to accessing the electrical or 

powered devices in the electrical enclosure. All components in the lower 

(wetted parts) section of the enclosure can be accessed for service or routine 

Figure 13 - Pressure regulator and T-strainer

6.1 Calibration

Calibration kit (Cat. No. 28144S) includes:
125 ml of zero solution (solution #1)
125 ml of calibration solution (solution #2) 1 each replacement cuvette
The SSR-Ex instrument was tested and calibrated prior to leaving the 

factory. The instrument operates from a pre-determined calibration 

curve for high accuracy of residual oxidant concentration. It is not 

necessary to recalibrate to maintain stated accuracy specifications. If 

re-calibration is required by a regulatory authority, calibration can easily 

be accomplished in the field using the procedure below.
Step 1:   Ensure the supply water is shut off. From the control buttons, 

enter the calibration screen. Follow the on-screen instructions.

Step 2:   Disconnect the black tubing from the pressure regulator to 

the pump. Some water may drain from the regulator, this is 

normal and it may be appropriate to use a container to catch 

the small volume of water.

Step 3:   Insert the black tube into solution #1. This is used to set a 

zero reference. Follow the instructions on the display.

Step 4:   When the SSR-Ex has completed the zero procedure you 

will be instructed to move the tube to solution #2. Follow the 

instructions on the display.

Step 5:   The SSR-Ex will take a reading of the solution. If this reads 

from 9-11 mg/l (PPM), the instrument is operating within 

the stated specifications and no further action is required. 

If the reading is outside of this range the SSR-Ex will allow 

you to make an adjustment to bring it into range. Follow the 

instructions displayed on the SSR-Ex.

Step 6:   Restore the black tube to the pressure regulator and turn 

the water supply back on. Check for leaks before closing the 

enclosure door.

maintenance, while the SSR-Ex instrument is powered, air supplied, and 

while the main enclosure door is open. Under normal operating conditions, 

it is recommended that the main enclosure door is closed. In cases where 

the SSR-Ex instrument is in an environment considered to be Zone 0 for a 

period of time, power and air to the instrument must be removed.
General Condition Check - Once a month check the general condition 

of the instrument in operation by opening the enclosure door and 

visually inspecting the instruments wet chemistry section components to 

confirm proper operation and to also check for any leaking components 

or other concerns. The wet section components can be accessed for 

maintenance, reagent replacement and visual inspection during voyages.

Specific condition of use: Parts of the enclosure are non-conducting 

and exceed the maximum permissible resistance according to 

the IEC 60079-0. Therefore, to avoid electrostatic charge build-

up, it must not be rubbed or cleaned with solvents or a dry cloth 

when installed/used within a potentially explosive atmosphere.

Содержание HF scientific SSR-Ex 28037

Страница 1: ...Model SSR Ex 28037 IOM HF SSR Ex 28037 WARNING Read this Manual BEFORE using this equipment Failure to read and follow all safety and use information can result in death serious personal injury property damage or damage to the equipment Keep this Manual for future reference IOM HF SSR Ex 28037 Rev 5 ...

Страница 2: ...verage Menu 13 5 2 1 8 Backlight Menu 13 5 2 1 9 Cycle Time Menu 14 5 2 1 10 Contact Period Menu 14 5 2 1 11 Standby Pulse Menu 14 5 2 1 12 Restore Defaults Menu 14 5 2 1 13 Restore Factory Cal Menu 14 5 2 1 14 About Device Menu 14 5 3 System Start Up 15 5 3 1 Flush Water Sample Line 15 5 3 2 Prepare Reagents 15 5 3 3 Remote Start 15 5 4 Typical Cycle 15 5 5 Observe Operation 15 6 0 Maintenance 15...

Страница 3: ...d the Manual This pictorial alerts you to electricity electrocution and shock hazards 1 0 Safety 2 0 Overview This instrument has been specifically designed for TRO measurement and to operate in hazardous atmospheres meeting the IECEx and ATEX ratings as stated in the specifications The instrument uses an IECEx approved purge pressurization controller called the Air Sentinel II Please refer to the...

Страница 4: ...ve FKM Polypropylene 316 Stainless Steel Acetal EPDM Silicone Modified PPO Thermoplastic Elastomer Sample Temperature Range 0 C 55 C 32 F 131 F Power Supply 100 to 230VAC 47 63 Hz 240VA Insulation Rating Double Insulated Pollution Degree 2 Overvoltage Category II Enclosure Powder coated steel with windows IP55 rating Environmental Conditions Altitude up to 2000 meters Up to 95 RH non condensing IE...

Страница 5: ...nside Front of Electrical Panel Enclosure Warning and informational Front of Unit Informational Front of Unit 2 3 Instrument Labels The following labels should be applied to the outside of the enclosure 3 0 Components Listing 3 1 Electronics Compartment WETTED COMPONENTS COMPARTMENT ACCESSIBLE ANY TIME CONTAINS REAGENT BOTTLES OPTICAL MEASUREMENT COMPONENTS INLET WATER SAMPLE PUMP PERISTALTIC CONN...

Страница 6: ...e for use with 6mm tubing if using the supplied fittings for installation ensure that a proper thread sealant is used Element Replacement Indicator The red ring will pop up to show when the element needs to be replaced Figure 7 Air Dryer Filter 4 0 Installation 4 1 Mounting and Site Selection The instrument is designed for wall mounting Choose a location that is easily accessible for operation and...

Страница 7: ...valve should be located close to the instrument to allow for periodic maintenance EMERGENCY DRAIN NO CONNECTION NEEDED SAMPLE WATER 6mm connection MAIN DRAIN 12 mm connection PURGE AIR AIR ONLY 6mm connection NOTE IMAGE DEPICTS BOTTOM OF THE ENCLOSURE DO NOT MIX UP THE AIR and WATER CONNECTIONS INSTALLING THE WATER INTO THE AIR CONNECTION WILL VOID THE WARRANTY Figure 10 Utility Connections All we...

Страница 8: ...n every bus may require a 120 ohm termination resistor to eliminate the possibilities of signal reflection on the line Do not run RS 485 cables in the same conduit as power Ensure each instrument is not powered when connecting the RS 485 line To prevent damage to the instrument ensure that power is disconnected prior to making connections The Modbus connection is located on the lower right portion...

Страница 9: ...he reading if True False 00007 Prime mode Start a prime if True False 4 5 2 Input Status These single bit values are readable from the master The data will be returned with the lowest addressed input status in the least significant bit LSB of the data Unused bits in the data will be set to 0 4 5 2 1 Valid Command s Code Name Broadcast 0x02 Read Input Status No 4 5 2 2 Format 16 bit word format MSB...

Страница 10: ... mantissa 4 5 4 3 Valid Addresses 30001 30XXX Address Type Register Function 30001 30002 Float Sensor read ing The meter reading 30003 30004 Float ABS The absorbance 30005 Int Main version Main board software version 30006 Int Display version Display board software version 30007 Int NOT USED 30008 Int Model number 28037 30009 Int Model suffix number Options model dependent 0 30010 Int Reading stat...

Страница 11: ...R Ex The SSR Ex is configured using the buttons located on the outside of the electrical enclosure The Mainscreen is shown below in Figure 12 Main Screen Using the UP and DOWN buttons navigate to the desired function shown on the screen Once you have highlighted the desired function press SELECT to display the next menu The Service and Config menus permit access to the processes listed in the char...

Страница 12: ...resentative sample to the SSR Ex Every time SERVICE mode is exited all faults are cleared If the original fault or a new fault occurs it will be posted to the display Faults will not stay cleared until the fault condition is no longer detected Select using UP DOWN buttons HOME No Change PRIME CALIBRATE WATER PRIME SEE BELOW CONFIG ANALOG OUTPUT see 4 2 1 3 CONFIG MODBUS see 4 2 1 4 ALARMS see 4 2 ...

Страница 13: ...S ANALOG OUTPUT MODBUS ALARMS 5 2 1 4 Modbus Menu 5 2 1 5 Alarms Menu The factory default setting is 9600 8 bit no parity 1 stop bit RTU All parameters are adjustable Make sure the parameters on the Host and the SSR Ex are configured identically Select using UP DOWN buttons These settings allow the user to configure the SSR Ex MODBUS Settings include the MODE BAUD RATE ADDRESS PARITY and BITS Sele...

Страница 14: ...n OFF Default setting is ON and default time is 24 Hr This option will restore the factory default settings Screen displays RESET DEFAULTS Pressing the SELECT button will restore factory default settings Before performing a Restore Defaults it is advised to make a listing of any user changed menu options prior to the reset 5 2 1 12 Restore Defaults Menu 5 2 1 11 Standby Pulse Menu RESTORE DEFAULTS...

Страница 15: ...e enclosure needs to be accessed for service ensure all power has been removed from a safe location prior to accessing the electrical or powered devices in the electrical enclosure All components in the lower wetted parts section of the enclosure can be accessed for service or routine Figure 13 Pressure regulator and T strainer 6 1 Calibration Calibration kit Cat No 28144S includes 125 ml of zero ...

Страница 16: ... when changing the reagents as they are corrosive These reagents may stain clothing After changing the reagents operators should wash their hands Step 1 Press SERVICE to stop the flow of sample water and drain the cuvette Step 2 Ensure the source water is turned off Step 3 Remove both reagent bottles by unscrewing cap assembly Step 4 The buffer is installed on the right and the indicator is instal...

Страница 17: ...olatile error Try factory reset Main Board Lost Internal communication loss Check cable connections try factory reset Low Setpoint Alarm Reading below set point Check source water oxidant level High Setpoint Alarm Reading above set point Check source water oxidant level 10 0 Spare Parts Listing Part Kit Name Description Frequency of part or kit change Calibration Kit 28144S Description Includes 12...

Страница 18: ..._ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ ...

Страница 19: ..._ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ ...

Страница 20: ... replacing other property which is damaged if this product does not work properly other costs resulting from labor charges delays vandalism negligence fouling caused by foreign material damage from adverse water conditions chemical or any other circumstances over which the Company has no control In addition the Company shall not be responsible for any costs incidental to the Company s warranty res...

Отзывы: