12
5.2.1 Menu Selection Options
5.2.1.1 Main Menu
SERVICE MENU
By design the SSR-Ex will only be actuated for
measurement by a remote command. The system can be set up
to either be actuated through MODBUS or by means of a remote
actuation provided by the end user.
PRIME function pulses the reagent pumps to ensure the reagent lines
are filled with reagent.
CALIBRATE function allows you to calibrate the SSR-Ex in the field.
Please refer to Section 5.1 - Calibration for the calibration procedure.
WATER PRIME function allows the sample pump to be actuated to
flush the water line to the SSR-Ex in order to expedite delivery of
representative sample to the SSR-Ex.
Every time SERVICE mode is exited, all faults are cleared. If the original
fault or a new fault occurs, it will be posted to the display. Faults will not
stay cleared until the fault condition is no longer detected.
(Select using UP & DOWN buttons)
HOME
No Change
PRIME
CALIBRATE
WATER PRIME
(SEE BELOW)
CONFIG
ANALOG OUTPUT
(see 4.2.1.3)
CONFIG
MODBUS
(see 4.2.1.4)
ALARMS
(see 4.2.1.5)
UNITS
(see 4.2.1.6)
AVERAGE
(see 4.2.1.7)
BACKLIGHT
(see 4.2.1.8)
CYCLE TIME
(see 4.2.1.9)
CONTACT PERIOD
(see 4.2.1.10)
STANDBY PULSE
(see 4.2.1.11)
RESTORE DEFAULTS
(see 4.2.1.12)
RESTORE FACTORY CAL
ABOUT DEVICE
(see 4.2.1.13)
(see 4.2.1.14)
SERVICE
PRESS SELECT TO
ACCESS
Once changes have been made to the settings, the SSR-Ex will display
the prompt “Are you sure?” To save the changed settings, press the
SELECT button.
5.2.1.2 Config Menu