14
(Select using UP & DOWN buttons) Values are in seconds.
Default value is 45 seconds.
The SSR-Ex will perform a test based on the cycle time interval
selected and actuation by means of MODBUS or RST.
(Select using UP & DOWN buttons)
The range 0:00 to 10:00 in minutes and seconds. The default is 00:00
This setting enables the user to allow contact time with the ballast
water and selected oxidant.
5.2.1.10 Contact Period Men
5.2.1.9 Cycle Time Menu
45 - 600
CYCLE TIME
0:00
CONTACT PERIOD
OFF - 100
STANDBY PULSE
(Select using UP & DOWN buttons)
This setting will energize the reagent pumps to ensure the lines remain
free of clogs. Range of settings is 1 - 100 hours. Setting below 1 turns
the function OFF.
Default setting is ON and default time is 24 Hr.
This option will restore the factory default settings. Screen displays
“RESET DEFAULTS?” Pressing the SELECT button will restore factory
default settings. Before performing a Restore Defaults it is advised to
make a listing of any user changed menu options prior to the reset.
5.2.1.12 Restore Defaults Menu
5.2.1.11 Standby Pulse Menu
RESTORE DEFAULTS
RESET DEFAULTS?
This option will restore the factory default calibration. Screen displays
“RESTORE FACTORY CAL?” Pressing the SELECT button will restore
the factory calibration. This will restore the factory reading, 4 and 20mA
calibrations.
5.2.1.13 Restore Factory Cal Menu
RESTORE FACTORY CAL
RESTORE FACTORY CAL?
(Select using UP & DOWN buttons)
This option will display the DISPLAY and MAIN software version
numbers.
5.2.1.14 About Devise Menu
DISPLAY: VERSION XX MAIN: VERSION:
XX
ABOUT DEVICE