4
323S
323U
323Du
400
400
400
220
220
220
Features
323E
323S
323U
323Du
Manual control
•
•
•
•
15-400 rpm 27:1 speed control
•
3-400 rpm 133:1 speed control
•
•
•
1.5-220 rpm 147:1 speed control
•
•
•
Auto-restart
•
•
•
Keypad lock
•
•
•
MemoDose
•
•
•
Analogue speed control
•
•
Remote stop/start control
•
•
Remote direction control
•
•
RS232 control
•
313D and 314D pumpheads
•
•
•
•
501RL pumphead
•
•
•
314MC and 318MC pumpheads
•
•
•
•
~100-120V/~220-240V operation
•
•
•
•
IP31 wipe down case
•
•
•
•
Switching the pump on
Switching the pump on
Switching the pump on
Switching the pump on
Switching the pump on
•
Switch on the mains power at the rear of the pump.
•
If the pump starts running look for the
!
symbol on the display. This
!
symbol indicates that the pump is set for
auto-restart. Press
key if you need to stop the pump.
•
The pump is now ready for manual control.
Auto-restart facility
Auto-restart facility
Auto-restart facility
Auto-restart facility
Auto-restart facility
Auto-restart will restart the pump after mains power interruptions. The pump will return to its previous operating
state. To install auto-restart:
•
Mains power must be available to the pump to engage auto-restart.
•
Stop the pump. Turn off the mains power switch at the rear of the pump.
•
Hold down the
key and turn the mains power switch on. The
!
symbol will show on
the display.
•
Start the pump. If the mains supply is interrupted the pump will automatically restart when the mains power
returns.
•
Auto-restart is retained while the pump is switched off.
•
To remove auto-restart switch off the mains power at the rear of the pump. Hold down the
key and turn the
mains power switch on. The
!
symbol will go out.
Do not use auto-restart for more than 10 starts per hour. We recommend remote control where
a high number of starts is required.
Содержание 323Du