Replacing the ESI probe sample capillary
5-67
To remove the existing capillary:
1. Remove the probe from the source (see
).
2. Use the combined 2.5-mm Allen wrench and cone extraction tool to
remove the 3 probe end cover retaining screws.
Warning:
The probe and source components can be contaminated
with biohazardous and/or toxic materials. Always wear
chemical-resistant, powder-free gloves while performing this
procedure.
Warning:
The probe and source can be hot. To avoid burn injuries, take
great care while performing this procedure.
Warning:
The ESI probe tip is sharp. To avoid puncture wounds, handle
the probe with care.
End cover retaining screws