Replacing the IntelliStart Fluidics tubing 5-113
8. Using a long “finger-tight” PEEK nut and 1/32-inch PEEK ferrule,
connect the red, 1/32-inch, 200-mm, PEEK tubing from port 4 of the
lock-spray selector valve to the left-hand side of the flow sensor, where
you use the 1/32-inch, 6-40 Valco™ compression fitting assembly.
Long “finger-tight” PEEK nut and 1/32-inch, PEEK ferrule:
1/32-inch, 6-40 Valco compression fitting assembly:
Caution:
The narrower, 1/32-inch tubing can slip out of the port.
Once the fitting is tightened, give the tubing a gentle tug to
ensure that it is gripped correctly.
A
B
C
A
B
C
A
B
C
Waters
A
B
C
Flow sensor
Содержание Xevo G2 QTof
Страница 10: ...x...
Страница 18: ...xviii Table of Contents...
Страница 34: ...1 16 Waters Xevo G2 QTof Overview...
Страница 44: ...2 10 Preparing the Mass Spectrometer for Operation...
Страница 108: ...5 32 Maintenance Procedures 2 Slide the collar to the end of the tool 3 Insert the collar in the sample cone Collar...
Страница 227: ...Connecting the liquid waste line B 17 Positioning of drain tube TP01807 Correct Incorrect...
Страница 238: ...B 28 External Connections...
Страница 242: ...C 4 Materials of Construction and Compatible Solvents...
Страница 248: ...Index 6...