
2-6
Preparing the Mass Spectrometer for Operation
To install the low-volume vials
:
1. If a standard reservoir bottle is fitted, remove it.
2. Screw the low-volume adaptors into the manifold and finger-tighten
them.
3. Screw the low-volume vials into the adaptors.
4. For each low-volume vial, ensure that the ends of the solvent delivery
tubes are positioned so that they are close to, but do not touch, the
bottom of the vial (see
Warning:
The reservoir bottles can be contaminated with
biohazardous and/or toxic materials. Always wear
chemical-resistant, powder-free gloves while performing this
procedure.
Warning:
Low-volume glass vials are fragile and can shatter,
cutting fingers. Take care when screwing them in, and never use
force.
Low-volume vial
Low-volume adaptor
Solvent delivery tube
Содержание Xevo G2 QTof
Страница 10: ...x...
Страница 18: ...xviii Table of Contents...
Страница 34: ...1 16 Waters Xevo G2 QTof Overview...
Страница 44: ...2 10 Preparing the Mass Spectrometer for Operation...
Страница 108: ...5 32 Maintenance Procedures 2 Slide the collar to the end of the tool 3 Insert the collar in the sample cone Collar...
Страница 227: ...Connecting the liquid waste line B 17 Positioning of drain tube TP01807 Correct Incorrect...
Страница 238: ...B 28 External Connections...
Страница 242: ...C 4 Materials of Construction and Compatible Solvents...
Страница 248: ...Index 6...