3-6
ESI and ESCi Modes of Operation
4.
Use the needle-nose pliers to remove the blanking plug from the corona
pin mounting contact.
Tip:
Store the blanking plug in a safe location.
Corona pin mounting contact:
5.
Use the needle-nose pliers to fit the corona pin to the mounting contact.
Requirement:
Ensure that the corona pin is securely mounted and that
its tip aligns with the sample cone orifice.
Warning:
The corona pin tip is sharp. To avoid puncture wounds,
handle the corona pin with care.
Caution:
To avoid damaging to the corona pin’s tip and bending
the pin, use the needle nose pliers to grip the corona pin at the
end that fits into the mounting contact.
TP02660
Corona pin mounting
contact blanking plug