Installing the corona pin
3-5
Installing the corona pin
Required materials
•
Chemical-resistant, powder-free gloves
•
Needle-nose pliers
•
80:20 acetonitrile/water
To install the corona pin:
1.
In the ACQUITY UPLC Console, click Standby
, and confirm that
the Operate indicator is not illuminated.
2.
Open the instrument’s access door.
3.
Unlatch the source enclosure door’s handle and open the door.
Warning:
The ACQUITY UPLC system connections, ESI probe,
and source can be contaminated with biohazardous and/or toxic
materials. Always wear chemical-resistant, powder-free gloves
while performing this procedure.
Warning:
To avoid electric shock, ensure that the instrument is in
Standby mode when commencing this procedure.
Warning:
The source can be hot. To avoid burn injuries, take
great care while working with the instrument’s access door open.
Warning:
The probe tip is sharp. To avoid puncture wounds, take
great care while working with the source enclosure door open if
an ESI probe is fitted.
Caution:
Do not apply any downward force to the source
enclosure door while the door is open.