Starting the TQ Detector
2-7
Running the TQ Detector at high flow rates
ACQUITY UPLC is run at high flow rates. To optimize desolvation, and thus
sensitivity, the ACQUITY TQD system should be run at appropriate gas flows
and desolvation temperatures. IntelliStart automatically sets these when you
enter a flow rate, according to the following table.
Note:
Under low ambient temperature, high moisture, and high flow rate
conditions, condensation can occur in the instrument’s source.
Flow rate versus temperature and gas flow :
Flow rate
(mL/min)
Source temp (°C)
Desolvation
temp (°C)
Desolvation gas
flow (L/hr)
0.000 to 0.100
120
250
500
0.101 to 0.300
120
350
600
0.301 to 0.500
150
400
800
>0.500
150
45
0
9
00