MODEL WPMZ-3-□□□-□□-□□□
10/12
IM-0881-04
15-3.
Wave compare
This is a mode to compare measurement waveform and
comparison waveform, output when alarming and log the
waveform.
Display [OK] when measurement waveform is
between comparison waveforms at all times. Display [NG] when
measurement waveform is out of comparison waveforms even
once and output alarm output. You can log measurement
waveform data each OK data and NG data up to 4 waveforms.
Parameter
1st Layer
2nd Layer
3rd Layer
1.Input
WaveA
WaveB
PatternSelect
StartCondition
Threshold
ThresholdOnDir
ThresholdTimeout
StartPosition
RefWaveCapture
CompWavePos
CreateCompWave
AutoScale
Scale
3.Display
DispSelect
WaveSelect
4.System
WaveLog
Overwrite
5.Diag/Log
OK/NG Wave
A/B View
OK 1
~
4
NG 1
~
4
OK/NG Wave
A/B Erase
OK 1
~
4
NG 1
~
4
For detail and how to use this mode, please refer to the
detailed manual which can be downloaded from our home page.
16.
INITIALIZE
Setting values can be initialized to factory default settings or
user saved settings.
16-1.
LOCATIONS OF PARAMETERS FOR INITIALIZE
1st Layer
2nd Layer
3rd Layer
4.System
Initialize
UserDefaultSave
UserDefaultLoad
FactoryDefaultLoad
*Refer to “
7. SETTING LAYER OVERVIEW
”.
16-2.
INITIALIZE TO THE FACTORY DEFAULTS
By performing the following steps, setting values can be
initialized to the factory defaults.
(Setting)→4.System→Initialize→FactoryDefaultLoad
Execute [OK]
Press Enter key
4
th
Layer
Message
Message
Message
3
rd
Layer
Back to
3rd Layer
Press Enter key to execute
Finish
16-3.
INITIALIZE TO USER DEFAULTS
Beside of initializing to factory defaults, current setting values
can be saved as user default data and setting values will be
initialized to the data.
16-3-1
.
Saving User Defaults
For initializing to user default values, entering setting values
used as default values are required.
(Setting)→4.System→Initialize→UserDefaultSave
1
)
Setting each values
Set all setting values which are needed as user default
values.
2
)
Saving user initial values
By executing to save user default values, current setting
values are saved as user default values.
Steps for the save operation are same as “
16-2. Initializing to
the factory defaults
”. Operate the WPMZ according to the
setting displays.
16-3-2
.
Initializing To User Defaults
(Setting)→4.System→Initialize→UserDefaultLoad
By executing to initialize to user defaults, current setting
values are replaced by user defaults.
Steps for the initialize operation are same as “
16-2. Initializing
to the factory defaults
”. Operate the WPMZ according to the
setting displays.
17.
KEY LOCK FUNCTION
To prevent unconsidered key operations, the WPMZ has the key
lock function.
Light when key lock is active
Press and hold over 1sec
→Key lock active/inactive
18.
ERROR CODES
When some malfunctions occur, error codes are displayed
according to the factor of the error.
ERROR
CODE
ERROR MESSAGE
RECOVERY PROCEDURE
E000
Program sum error
During the error mode, hold
down the ENTER key for 1
second (long-press) to reset or
power down and on.
*If the WPMZ does not
recover by this procedure,
please contact your dealer or
our company.
E006
RAM error
E100 to
102
Errors associated with
serial flash memory
E103 to
105
E210 to
211
Errors associated with
FRAM
E202 to
203
Errors associated with
calibration values
E110 to 111
Error associated with
sensor power short
E204 to
205
Errors associated with
setting values
E206 to
209
Errors associated with
initial values
Other than
above
codes
Other errors
・
If start-up delay is enabled, the WPMZ displays “-----” according
to the delay time.
・
If display value becomes out of displayable range, “OVER” is
displayed in the display.
ENTER
DISP
MENU
FUNC
Indicates KEY LOCK is valid
KEY LOCK on:
Enabled by pushing DISP
key and ENTER key
simultaneously
* lights up.
KEY LOCK off:
Disabled by pushing DISP
key and ENTER key
simultaneously
* lights down.