WATANABE WPMZ-3 Series Скачать руководство пользователя страница 10

 

 

MODEL WPMZ-3-□□□-□□-□□□                                                                                 

 

10/12 

 

                                                                                                                                                                                                               

IM-0881-04 

15-3. 

 

Wave compare 

This  is  a  mode  to  compare  measurement  waveform  and 

comparison  waveform,  output  when  alarming  and  log  the 
waveform.

 

Display  [OK]  when  measurement  waveform  is 

between comparison waveforms at all times. Display [NG] when 
measurement  waveform  is  out  of  comparison  waveforms  even 
once  and  output  alarm  output.  You  can  log  measurement 
waveform data each OK data and NG data up to 4 waveforms. 
 
Parameter 

1st Layer 

2nd Layer 

3rd Layer 

1.Input 

WaveA 
WaveB 

PatternSelect 
StartCondition 
Threshold 
ThresholdOnDir 
ThresholdTimeout 
StartPosition 
RefWaveCapture 
CompWavePos 
CreateCompWave 
AutoScale 
Scale 

3.Display 

DispSelect 

WaveSelect 

4.System 

WaveLog 

Overwrite 

5.Diag/Log 

OK/NG Wave 

A/B View 

OK 1

NG 1

OK/NG Wave 

A/B Erase 

OK 1

NG 1

For  detail  and  how  to  use  this  mode,  please  refer  to  the 

detailed manual which can be downloaded from our home page. 

 

16. 

 

INITIALIZE 

Setting  values  can  be  initialized  to  factory  default  settings  or 
user saved settings.   

16-1. 

 

LOCATIONS OF PARAMETERS FOR INITIALIZE 

1st Layer 

2nd Layer 

3rd Layer 

4.System 

Initialize 

UserDefaultSave 

UserDefaultLoad 

FactoryDefaultLoad 

*Refer to “

7. SETTING LAYER OVERVIEW

”. 

 

16-2. 

 

INITIALIZE TO THE FACTORY DEFAULTS 

By performing the following steps, setting values can be 
initialized to the factory defaults.

 

 

 
(Setting)→4.System→Initialize→FactoryDefaultLoad 

Execute [OK]

Press Enter key

4

th

Layer

Message

Message

Message

3

rd

Layer

Back to

3rd Layer

Press Enter key to execute

Finish

 

 
 

16-3. 

 

INITIALIZE TO USER DEFAULTS 

Beside of initializing to factory defaults, current setting values 
can  be  saved  as  user  default  data  and  setting  values  will  be 
initialized to the data.

 

16-3-1

 

Saving User Defaults 

For  initializing  to  user  default  values,  entering  setting  values 
used as default values are required. 
  (Setting)→4.System→Initialize→UserDefaultSave 

1

 

Setting each values 
Set  all  setting  values  which  are  needed  as  user  default 
values. 

2

 

Saving user initial values 
By  executing  to  save  user  default  values,  current  setting 
values are saved as user default values. 
Steps for the save operation are same as “

16-2. Initializing to 

the  factory  defaults

”.  Operate  the  WPMZ  according  to  the 

setting displays. 

 

16-3-2

 

Initializing To User Defaults 

(Setting)→4.System→Initialize→UserDefaultLoad 

By  executing  to  initialize  to  user  defaults,  current  setting 
values are replaced by user defaults. 
Steps for the initialize operation are same as “

16-2. Initializing 

to  the  factory  defaults

”.  Operate  the  WPMZ  according  to  the 

setting displays.

 

 
 

17. 

 

KEY LOCK FUNCTION 

To prevent unconsidered key operations, the WPMZ has the key 
lock function. 

   

Light when key lock is active

Press and hold over 1sec
→Key lock active/inactive

 

 

18. 

 

ERROR CODES 

When  some  malfunctions  occur,  error  codes  are  displayed 
according to the factor of the error. 

 
 
 
 
 
 
 

 

ERROR 

CODE 

ERROR MESSAGE 

RECOVERY PROCEDURE 

E000 

Program sum error 

During the error mode, hold 
down the ENTER key for 1 
second (long-press) to reset or 
power down and on. 
 
*If the WPMZ does not 
recover by this procedure, 
please contact your dealer or 
our company. 

E006 

RAM error 

E100 to 

102 

Errors associated with 
serial flash memory 

E103 to 

105 

E210 to 

211 

Errors associated with 
FRAM 

E202 to 

203 

Errors associated with 
calibration values 

E110 to 111 

Error associated with 
sensor power short 

E204 to 

205 

Errors associated with 
setting values 

E206 to 

209 

Errors associated with 
initial values 

Other than 

above 

codes 

Other errors 

 

If start-up delay is enabled, the WPMZ displays “-----” according 

to the delay time. 

If display value becomes out of displayable range, “OVER” is 

displayed in the display. 

 
 
 
 

 

ENTER

DISP

MENU

FUNC

Indicates KEY LOCK is valid

KEY LOCK on:

Enabled by pushing DISP 
key and ENTER key 
simultaneously
*   lights up.

KEY LOCK off:

Disabled by pushing DISP 
key and ENTER key 
simultaneously
*    lights down.

Содержание WPMZ-3 Series

Страница 1: ...RTU Input Ach Input Bch GO output GO output 2 Do not wire the power supply line input signal lines and output signal lines near noise sources or relay drive lines 3 Bundling or containing in a same d...

Страница 2: ...ocess input Suitable wire AWG 24 to 16 GO OUT GO GND V HI LO EXC EXC A HI 20 14 15 16 17 18 19 27 21 22 23 24 25 26 connected internally connected internally 22 25 Process input GO output Ach Bch No N...

Страница 3: ...iffelenceHold AVE AveHold Hold mode Indicate hold mode with icon In addition 1ch product shows abbreviation below NORM NormalHold AREA AreaHold 3rd item Displays measured value of 3rd item 2nd item Di...

Страница 4: ...ve 32 None 2 4 8 16 32 64times Set moving average of input Offset 0 Go to screen of setting offset Only process input Fullscale 10000 Go to screen of setting fullscale Only process input ManuAdjust 0...

Страница 5: ...V 10V 1 5V 0 20mA 4 20mA Select a type of output range OutputDispValue None None Ach Bch Calc Select a ch referenced for analog output OutputScale 0 10000 0 99999 100 99999 Scaling of analog output Se...

Страница 6: ...lear Erase all alarm log 1st Layer 2nd Layer 3rd Layer setting items 4th Layer setting values Remarks Initial values Initial values 1 Input Wave Compare A Wave Compare B PatternSelect Pattern in use P...

Страница 7: ...r DispValue manually that you want to display Back to one before by press left arrow key Go to Fullscale display Near 100 calibration Operation is same as Settings are saved by pressing MENU key and b...

Страница 8: ...threshold value to judge How to set the value is shown below The following is an explanation for the compare mode of Level judgement Setting 2 Output CompareAL1 Threshold Set value with arrow keys Pre...

Страница 9: ...and performs same action of DISP key TREND LOG FUNCTION Trend log function is to acquire the alarm log any time by external control It is a function to execute with While data is being acquired the tr...

Страница 10: ...System Initialize UserDefaultSave 1 Setting each values Set all setting values which are needed as user default values 2 Saving user initial values By executing to save user default values current set...

Страница 11: ...etween Power terminals and inputs external controls comparative outputs other outputs AC1500V for 1 minute AC power and DC power Between input terminals and external controls comparative outputs other...

Страница 12: ...upper limit or lower limit value AL3 Display value AL4 upper limit or lower limit value AL4 Difference judgement mode When the maximum minimum during the fixed time exceeds the judgement value alarm...

Отзывы: