32
Sterilization and storage
W&H recommends sterilization according to EN 13060, class B .
> Note the instructions of the unit manufacturer .
> Clean, disinfect and lubricate prior to sterilization .
> Wrap the screwdriver angled and accessories in sterile goods packaging according to EN 868-5 .
> Make sure that you only remove dry sterile goods from the sterilizer .
> Store sterile goods in a dust-free and dry condition .
The standard ISO 7785-2/ISO 14457 stipulates a durability of at least 250 sterilization cycles .
In the case of the W&H screwdriver angled, we recommend you carry out a regular service after 1000
sterilization cycles or one year .