54
Technical data
S-N2 (REF 30285000) is waterproof according to IPX8, 1 m depth of immersion,
1 hour (water-tight in accordance with IEC 60529)
Pollution level:
2
Overvoltage category:
II
Altitude:
up to 3,000 m above sea level
Class II medical electrical equipment (protective earth conductor used for functional earth connection only!)
Type B applied part (not suitable for intracardiac application)
S-N2 (REF 30285000) is approved for operation in potentially explosive atmospheres.
Classification according to Paragraph 6 of the General Specifications for the Safety of Medical Electrical Equipment
according to IEC 60601-1/ANSI/AAMI ES 60601-1
AP