41
Hygiene and maintenance
Motor with cable / Universal support / Irrigant support
> Do not immerse the motor with cable, the universal support or the irrigant support in liquid disinfectant or in an
ultrasonic bath.
Motor with cable / Universal support / Irrigant support
> Clean the motor with cable, the universal support and the irrigant support under running tap water (< 35°C / < 95°F).
> Rinse and brush off all internal and external surfaces.
> Remove any liquid residues using compressed air.
Control unit
> Do not immerse the control unit in water or clean under running water.
Foot control
> The ESD spring contact on the bottom of the foot control must be cleaned regularly.
Manual cleaning