Hygiene and maintenance
Motor with cable / Universal support / Irrigant support
W&H recommends automated cleaning and disinfection using a washer-disinfector (WD).
Read the notes, follow the instructions and heed the warnings provided by the manufacturers of
washer-disinfectors, cleaning agents and/or disinfectants.
> The control unit and foot control are not approved for automated cleaning and disinfection.
Evidence of the basic suitability of the motor with cable, the universal support and the irrigant support for effective
automated disinfection was provided by an independent test laboratory using the »Miele PG 8582 CD« washer-disinfector
(Miele & Cie. KG, Gütersloh) and the »Dr. Weigert neodisher® MediClean forte« cleaning agent (Dr. Weigert GmbH & Co. KG,
Hamburg).
> Cleaning at 55 °C (131°F) – 5 minutes
> Disinfection at 93 °C (200°F) – 5 minutes
Automated cleaning and disinfection