Maintenance
78
Lara | Instructions for Use | RIL-1 W&H Med ENG Rev06 | 14/04/2022 | © W&H Sterilization Srl
5
Insert the new seal and press first up and down.
6
Press left and right and then the entire seal circumference to ensure its perfect
placement.
Notice
: A steam discharge can damage the sterilizer plastic parts. Ensure the seal
does not stick out.
7
Wipe any residual water and run a Vacuum and Helix test to check for perfect
tightness of the seal. See "Vacuum test" on page 46 and "Helix test" on page 44.