Maintenance
REPLACE THE CONSUMABLE BEFORE THE MAINTENANCE DUE
DATE
If you replace the consumables before the request of replacement
appears, you should manually reset the counters through the
following procedure.
1
On the homepage, tap
>
2
Select the consumable you want to replace: a message appears
showing the current worked hours of the part.
3
Tap
to see an animated replacement procedure.
4
When you have replaced the consumable tap
to confirm: the
executed cycle counter is reset.