Maintenance
74
Lara | Instructions for Use | RIL-1 W&H Med ENG Rev06 | 14/04/2022 | © W&H Sterilization Srl
CLEAN THE WATER TANKS
Notice
: Do not touch the water level sensors. If misplaced or misaligned from their original position, the operation of the sterilizer could be
impaired.
Proceed as follows:
1
Clean the internal tank surfaces with a soft sponge moistened with water. For
better cleaning results, clean with W&H MC-1000 cleaning solution. Use a small
non-abrasive brush to clean the areas that are difficult to reach.
Notice
: never use any other disinfectant, detergent or abrasive product, as they
might result aggressive for the tank material.
2
Rinse the internal tank surfaces throughly, until all residuals of dirt and detergent
have been eliminated.
3
Dry the internal tank surfaces.
4
Detach the drain tube, attach it to the drain port of the other tank and repeat steps
1, 2, 3.