Sterilizer tests
Lara | Instructions for Use | RIL-1 W&H Med ENG Rev06 | 14/04/2022 | © W&H Sterilization Srl
45
4
Place the tube with the capsule in the center of a tray in the lowest rack position
and close the chamber door.
5
On the homepage, tap
Helix-B&D test
.
6
To set the duration of the Plateau/Sterilization phase and other settings, tap
.
7
Tap and enter your credentials if required: the chamber door locks.
8
Wait until the end of the test and tap
OPEN
: the chamber door unlocks.
9
Enter your credentials if required.
10
Open the chamber door, extract the tray using the tray holder and take the tube.
CAUTION!
Risk of burns. The test pack is very hot at the end of the cycle. Wear
appropriate PPE (e.g. gloves).
11
Unscrew the tube capsule and remove the indicator strip.
12
Check the change in colour. See "Interpret the test result" on the next page