Chapter 7
Touchscreen Operation
82
actinic illumination without preceding saturation pulse analysis
(Initial pulse). Width of actinic illumination is adjusted as de-
scribed above for clock interval and initial pulse is selected by
the SET command. Further, actinic light intensity is adjusted in
the window “Light Sources” (Fig. 41, page 77).
An A Yield routine can be started using the START but-
ton on the side panel of Fig. 45. In this case, the screen display
will automatically switch to the A Yield experimental win-
dow (Fig. 32, page 60).
A Yield Settings
START
SET
EXIT
Side Panel
START
. Start Actinic
light + Yield routine and
switch to Actinic Light
and Yield window (Fig.
,
. Up/Down
keys to move to a menu
line.
SET
. Select item.
Width (mm:ss)
2:00
Initial Pulse
yes
Fig. 45:
A Yield Settings
Induction Curve
7.3.3.2
Induction curve experiments are configured in the window “In-
duction Curve Settings” (Fig. 46, page 83).
Delay (range 5 s to 10 min) defines the dark interval between
saturation pulse analysis with the dark-acclimated sample (F
0
,
F
M
determinations, Section 5.1.1, page 32) and beginning of ac-
tinic illumination.