Specifications
Chapter 8
105
8 Specifications MINI-PAM-II
8.1
Basic System
8.1.1 Optoelectronic Unit
MINI-PAM-II/B (Blue Version)
Measuring light:
Blue (470 nm) LED, standard modulation fre-
quencies 5 to 25 Hz, adjustable in increments of 5 Hz, and 100
Hz, measuring light PAR at standard settings = 0.05 µmol m
-2
s
-
1
. Fluorescence at wavelengths greater than 630 nm is measured
Actinic light:
Same blue LED as for measuring light, maximum
actinic PAR = 3000 µmol m
-2
s
-1
, maximum PAR of saturation
pulses = 6000 µmol m
-2
s
-1
adjustable at increments of 500
µmol m
-2
s
-1
.
MINI-PAM-II/R (Red Version)
Measuring light:
Red (655 nm) LED, modulation frequencies
and PAR as described for MINI-PAM-II/B. Fluorescence at wave-
lengths greater than 700 nm is measured
Actinic light:
Same red LED as for measuring light, maximum
PAR of actinic light and saturation pulses as described for MINI-
PAM-II/B
MINI-PAM-II/B and MINI-PAM-II/R
Far red light:
Peak emission at 735 nm
Signal detection:
PIN photodiode protected by long-pass and a
short-pass filters