Chapter 7
Touchscreen Operation
60
show a vertical cursor line as well as the x-y data of the intersec-
tion between cursor line and spectrum. Use the arrow keys to
navigate through the spectra. Touching the graph display again
returns to the original function of arrow keys.
7.2.6 A Yield
The “A Yield” window is the first of four windows designed
for starting and monitoring automated measuring routines. The
parameters of the A Yield routine can be set in the menu
“Program/Clock” (Fig. 43, page 80).
A Yield
A
START
MENU
Side Panel
START
. Start A Yield
routine.
Previous window (Ft-
Chart).
Next window (Induction
Curve).
MENU
. Go to Main Menu.
SAT #1 SAT #2
F
571
1293
Fm’
2065
1356
PAR
0
856
Y(II)
0.723
0.046
ETR
0.0
16.5
2018-04-12
8:20:09
Main Panel
SAT #1, SAT #2.
Column for data of
first and second saturation pulse analy-
sis, respectively.
F, Fm’
. Fluorescence level before and
at saturation pulse. For dark-acclimated
samples: F=F
0
and F
M
’=F
M
.
Y(II)
. Effective photochemical quantum
yield of PS II (F
V
/F
M
of dark-acclimated
samples).
PAR, ETR
. Photosynthetically active
radiation (µmol·photons m
-2
·s
-1
), elec-
tron transport rate
(µmol·electrons m
-2
·s
-1
).
Fig. 32: