Keep decreasing the supply voltage until 5.5 volts is reached, as measured
with a digital DVM. The meter should maintain correct operation even though
the small “L” or “Lo Bat” is on, indicating a low battery.

Now raise the voltage back up until the small “L” or “Lo Bat” turns off. The
voltage at which this happens should be from 5.50 to 5.55 volts.

Under normal operation, an additional 20 minutes will elapse from the time
the small “L” or “Lo Bat” appears until the battery voltage decreases to 5.5
volts, assuming a normal 5-pack of AA NiCad batteries is being used. Below
5.5 volts, the computer may introduce some errors or not operate the display
at all.

6.4  Zero Adjustment Using Wahl Calibrator

1.  Insert the Wahl calibrator and adjust the calibrator to the

32°F/0°C position. Apply power to the meter and make an initial adjustment
at R5 for a reading of 32.0°F or 0.0°C depending on your model.

NOTE: For the version 392HP, which has a 4-wire probe configuration, the
microprocessor averaging process is slightly slower. The moving average for
the 392HP may be made faster by turning the power off and then back on
again, any time an input resistance or pot setting is changed.

2.  Make the best adjustment you can at this point, and then proceed

to the span adjustment.

6.5  Span Adjustment Using Wahl Calibrator

1.  Set Wahl Calibrator for 284°F/140°C and apply power to the

meter. Adjust R35 (R16 on HP.) clockwise for a display of 284.0°F or
140.0°C.

2.  Repeat the zero span adjustment until the meter reads

32.0°F/0.0°C exactly, and reads 212°F/100°C or 284°F/140°C within accept-
able limits of error.
Since only one span adjustment covers both low and high scales, any error
that must be tolerated should be pushed to the high scale where percentage
wise it will be less of an error.

6.6  Adjustment Using Decade Resistance Box
Basic Theory: Platinum has a well defined resistance versus temperature
curve. Accordingly, if a decade resistor is put in the place of the probe to
simulate the standard platinum element, a quick and accurate calibration can
be achieved.

6.6.1.  Setup and Calculations
The decade resistor box may be used as the source to perform a standard
calibration. A coiled cord cable with connector for the Wahl probe should be
used (Model Number 10434-01). This has an approximate resistance of 0.36
ohms. The actual resistance should be determined for your setup and this
value used in the calculations. The total resistance to be simulated is .94
ohms above the sensor resistance specified in the ITS90 Platinum Resistance
Tables for 100 ohm, .00385 ohm/ohm/degree C sensors. Therefore, the
decade resistance setting should be a table value plus 0.94 ohms minus the
actual cable resistance measured above i.e. the decade box setting for 0°C
should be: 100.00 + 0.94 - 0.36 = 100.58 ohms.

Note: for HP models do not subtract cord resistance in calculation steps. Use
coiled cord Model Number 10434-02.

Step 1: Set the Decade Box for a total resistance (decade box + lead

resistance) for 100.94 ohms and adjust zero control R5 for a display reading
of 0.0°C or 32.0°F, ±0.1°.

Step 2: Set the Decade Box for a total resistance (decade box + lead

resistance) for 154.52 ohms and adjust span control R35 (R16 for HP mod-
els) for a display reading of 140.0°C or 284.0°F, ±0.1°.

Step 3: Repeat Steps 1 and 2 as necessary for optimum calibration.

6.7  Peak Hold Circuit

1.  To enable the PEAK HOLD mode, depress the forward/ upper

position of the Power Switch. Adjust calibrator or decade box for an upscale
temperature reading allowing the display to stabilize.

2.  When display has stabilized to a temperature reading appropriate

for the calibrator/decade box setting, move the calibrator/decade box to the
position or resistance that represents a lower temperature, and verify that the
peak temperature of Step 1 is held.

NOTE: This is a digital peak that is computed by the peak hold circuit, and
as such will not drift up or down as may happen with analog circuits. What
may happen is that the calibrator or decade box can momentarily “go open”
as it is switched between settings. The meter may see this and jump up scale
and retain a higher reading. When this happens, switch meter off and back
on to peak hold setting, and the reading will be restored.

Содержание HEAT-Prober 392HP

Страница 1: ...rober PLATINUM RTD THERMOMETERS Models 392HP 392E 392E S 392EX 392EX S 392EVX 392M 392M S 392MX 392MX S 392VX 360X and 700M INSTRUCTION MANUAL Copyright 2004 Wahl Instruments Inc Printed in USA W 392...

Страница 2: ...authorized to assume for Wahl any liablility except as set forth above Warranty Temperature Probes and Meter Accessories Wahl warrants its Hand Held Meter Temperature Probes to be free of defects in...

Страница 3: ...in all light conditions To minimize power drain the display is multiplexed The LCD display is 0 4 high with excellent contrast A large portion of the circuit is designed with low power C MOS componen...

Страница 4: ...to conserve battery life 2 1 Meter Operation The HEAT Prober instrument has an accurate measuring range as stated on the front label The probes however may not be capable of this entire range of use...

Страница 5: ...ble of 50 hours continuous operation 2 4 1 Recharging To recharge the batteries use the recharger supplied Plug in directly to 115VAC 60 cycle wall plug and insert jack into socket on left side of HEA...

Страница 6: ...body temperature it requires good contact with that body or in the case of liquids or gases good immersion and mixing A portable probe temperature sensor must have the following characteristics a a l...

Страница 7: ...hane Propane E Metal Dust F Coal Dust and G Grain Dust All 392 Models Available in F C or Dual Scale add F C or D to Model Number Dual Scale not available on S Models 392 SERIES METER SPECIFICATIONS M...

Страница 8: ...gure 5 1 392M Platinum Heat Prober with Wahl Calibrator Model Number CAL392HT 6 0 SET UP AND CALIBRATION 6 1 Equipment Needed A Digital voltmeter Fluke 41 2 digit Model 8600A or equivalent B Wahl Plat...

Страница 9: ...a quick and accurate calibration can be achieved 6 6 1 Setup and Calculations The decade resistor box may be used as the source to perform a standard calibration A coiled cord cable with connector fo...

Страница 10: ...nd 392EVX Models 392HP 392M and 392E Use the Circuit Board Diagrams on these pages to locate potentiometer placement of Zero Adjustment and Span Adjustment Note that placement varies with Model See pa...

Страница 11: ...entire temperature range This also results in superior system accuracies at all temperatures The HP option utilizes a Noise Rejection program which minimizes transient noise effects In this version a...

Страница 12: ...he program Details of the components will be discussed in this section 9 2 Program Flow When Power is turned on the program will start executing instructions from address 000 the location of the top l...

Страница 13: ...age be crammed with new data Additionally if over range is displayed indicating on open probe lead or removed probe conditions are set for the next entry into Gaincheck to cause the compensation readi...

Страница 14: ...58MHz square wave with levels from 0 to 5 volts Replace U6 If still no display go to Step 9 Replace U4 If still no dis play go to Step 7 4 Check voltage at Q4 collector It should be equal to the batte...

Страница 15: ...Return unit to factory for repair NOTE Does not apply to 392MX and 392M with Serial Numbers before N120 11 0 ACCESSORIES 392 SERIES CALIBRATORS Model Number Usable With Meter Calibration Points CAL39...

Страница 16: ...the same as the 392M which is a recharge able nickel cadmium battery pack composed of five AA size cells with a com bined voltage of 6 0 volts and capacity rating of 65 amp hours The unit is supplied...

Страница 17: ...0 plug connector 3 Set resistance to 29 490 ohms and adjust R5 for a display read ing of 0 00 C 32 F 4 Set the decade box to a resistance of 2760 ohms and adjust R13 for a display reading of 60 00 C 1...

Страница 18: ...to next step 8 Measure voltage at pin 6 of wire harness It should be equal to the battery or power supply voltage Proceed to next step Check wire harness for broken wires 9 Replace Q1 and repeat Step...

Страница 19: ...follow up reminders about recalibration dates 16 0 RETURN FOR REPAIR CALIBRATION SERVICE Warranty Repair Information Please contact Wahl Customer Service at 800 421 2853 for a Return Merchandise Auth...

Отзывы: