5.0  MAINTENANCE/REPAIR
The HEAT-Prober instrument may be recalibrated, or adjusted for a particular
probe in the field by following the directions in Section 6.0. However, it is
recommended that calibrations and repair be done at the factory. Return the
HEAT-Prober in its foam package with a description of symptoms, and a
repair estimate will be made for your approval.

The best way to check the functioning and calibration condition of the HEAT-
Prober is to use the Wahl accessory Calibrator Models Numbers CAL392HT
or CAL392LT, which plug right into the probe connection and provides five
key calibration points. (See Figure 5-1). The calibrator has a set of precision-
wound resistors which are measured at the factory against a NIST traceable
standard. The WAHL calibrator is a convenient instrument to determine
whether a problem is in the probe or the meter. If the meter checks on all 5
calibration points within tolerance, for example, then one will know that the
probes are at fault.

The most common maintenance problem
is battery failure, evidenced by no dis-
play. NiCad battery failure and prevention
of this problem are noted in Section 2.4.
If you suspect a dead battery:

1.  Plug in the recharger and

allow to recharge for 12 hours. If batter-
ies fail to recharge after this time:

2.  Open the battery compart-

ment lid on the bottom of the instrument.
Unplug the battery. Using a DVM, check
voltage with a 16 ohm load. If the voltage
drops rapidly it indicates a fault in the
battery.

3.  With the convenient detach-

able plug, the battery pack Model
Number 11681-1 or 11681-2 can be easi-
ly replaced by a new one. For emergen-
cies it is possible to use a 9-volt transis-
tor battery to power the HEAT-Prober (but
one must make a special connection).
Expected continuous running time on 9-
volt alkaline battery is about 3 hours, for
LED display or 6 hours for LCD display.

Figure 5-1. 392M Platinum Heat

Prober with Wahl Calibrator

Model Number CAL392HT.

6.0  SET-UP AND CALIBRATION

6.1  Equipment Needed

A.  Digital voltmeter, Fluke 41/2 digit Model 8600A or equivalent.
B.  Wahl Platinum Calibrator (Model Number CAL392HT or CAL392LT) or

General Radio decade box, Model 1433-1 (or equivalent) with standard coil
cord (Model 10434-01 or 10434-02) cable as used with Wahl probes.

C.  Power supply — Powermate, Model OPA-1OC or equivalent.

6.2  Voltage Regulator Adjustment (Note: Meters with serial numbers S/N
3445 and higher, or N Series,do not require adjustment of voltage regulator.)

Remove the 4 screws from the bottom of the case, then remove the top
cover. You will note the adjustment points as shown in Figure 6-1.

1.  Initially the 392M should be set-up with an external power supply

(rather than a battery) so that the current can be limited to no more than 120
milliamperes.

2.  Connect the plus (+) DVM lead to C-10 junction common with Q2.

Connect the minus (-) lead to the cathode lead of CR1.

3.  Apply power to the 392M using an external power supply with the

voltage set at 6.5V and current limit set at 120 milliamperes. Connect power
supply (+) to U1-1 and (-) to CR1 - cathode. Adjust R21 for voltage reading
of e5.000 volts on the DVM.

4.  When the 392M has been mounted in its case and connected to a

5-battery pack, the DVM is connected as described above and the +5.000
adjustment checked and trimmed if necessary.

6.3  Low Battery Adjustment and Testing

1.  Connect DVM (-) to CR1 Cathode and (+) to upper side of R24.
2.  Adjust R37 clockwise or counterclockwise until DVM reads 2.11 volts.
3.  Disconnect DVM.
4.  DISCONNECT BATTERY PACK. Low Battery indicator must be per-

formed with an external variable power supply. Start with 6 volts and slowly
decrease the voltage toward 5.8 volts.

When the battery threshold is reached, a small “L” should appear in the most
significant digit position of the 392HP, 392E and 392M display. “Lo Bat”
appears in the display of the 392EX, 392MX, 392EVX and 392MVX. The actu-
al value of the low battery voltage threshold may vary from 5.45 volts to 5.55
volts.

Содержание HEAT-Prober 392HP

Страница 1: ...rober PLATINUM RTD THERMOMETERS Models 392HP 392E 392E S 392EX 392EX S 392EVX 392M 392M S 392MX 392MX S 392VX 360X and 700M INSTRUCTION MANUAL Copyright 2004 Wahl Instruments Inc Printed in USA W 392...

Страница 2: ...authorized to assume for Wahl any liablility except as set forth above Warranty Temperature Probes and Meter Accessories Wahl warrants its Hand Held Meter Temperature Probes to be free of defects in...

Страница 3: ...in all light conditions To minimize power drain the display is multiplexed The LCD display is 0 4 high with excellent contrast A large portion of the circuit is designed with low power C MOS componen...

Страница 4: ...to conserve battery life 2 1 Meter Operation The HEAT Prober instrument has an accurate measuring range as stated on the front label The probes however may not be capable of this entire range of use...

Страница 5: ...ble of 50 hours continuous operation 2 4 1 Recharging To recharge the batteries use the recharger supplied Plug in directly to 115VAC 60 cycle wall plug and insert jack into socket on left side of HEA...

Страница 6: ...body temperature it requires good contact with that body or in the case of liquids or gases good immersion and mixing A portable probe temperature sensor must have the following characteristics a a l...

Страница 7: ...hane Propane E Metal Dust F Coal Dust and G Grain Dust All 392 Models Available in F C or Dual Scale add F C or D to Model Number Dual Scale not available on S Models 392 SERIES METER SPECIFICATIONS M...

Страница 8: ...gure 5 1 392M Platinum Heat Prober with Wahl Calibrator Model Number CAL392HT 6 0 SET UP AND CALIBRATION 6 1 Equipment Needed A Digital voltmeter Fluke 41 2 digit Model 8600A or equivalent B Wahl Plat...

Страница 9: ...a quick and accurate calibration can be achieved 6 6 1 Setup and Calculations The decade resistor box may be used as the source to perform a standard calibration A coiled cord cable with connector fo...

Страница 10: ...nd 392EVX Models 392HP 392M and 392E Use the Circuit Board Diagrams on these pages to locate potentiometer placement of Zero Adjustment and Span Adjustment Note that placement varies with Model See pa...

Страница 11: ...entire temperature range This also results in superior system accuracies at all temperatures The HP option utilizes a Noise Rejection program which minimizes transient noise effects In this version a...

Страница 12: ...he program Details of the components will be discussed in this section 9 2 Program Flow When Power is turned on the program will start executing instructions from address 000 the location of the top l...

Страница 13: ...age be crammed with new data Additionally if over range is displayed indicating on open probe lead or removed probe conditions are set for the next entry into Gaincheck to cause the compensation readi...

Страница 14: ...58MHz square wave with levels from 0 to 5 volts Replace U6 If still no display go to Step 9 Replace U4 If still no dis play go to Step 7 4 Check voltage at Q4 collector It should be equal to the batte...

Страница 15: ...Return unit to factory for repair NOTE Does not apply to 392MX and 392M with Serial Numbers before N120 11 0 ACCESSORIES 392 SERIES CALIBRATORS Model Number Usable With Meter Calibration Points CAL39...

Страница 16: ...the same as the 392M which is a recharge able nickel cadmium battery pack composed of five AA size cells with a com bined voltage of 6 0 volts and capacity rating of 65 amp hours The unit is supplied...

Страница 17: ...0 plug connector 3 Set resistance to 29 490 ohms and adjust R5 for a display read ing of 0 00 C 32 F 4 Set the decade box to a resistance of 2760 ohms and adjust R13 for a display reading of 60 00 C 1...

Страница 18: ...to next step 8 Measure voltage at pin 6 of wire harness It should be equal to the battery or power supply voltage Proceed to next step Check wire harness for broken wires 9 Replace Q1 and repeat Step...