Wahl HEAT-Prober 392HP Скачать руководство пользователя страница 7

8

9

392 S

ERIES

M

ETER

S

PECIFICATIONS

Model

392HP

392E 

392EX

392E-S
392EX-S

392EVX

Range

-290 to 1450°F

-180 to 788°C

-290 to 1450°F

-180 to 788°C

-290 to 1450°F

-180 to 788°C

Digital Display

Bright 0.33"

LED

392E/392E-S

Bright 0.33" high LED

Red, 0.4" high

LCD

392EX/392EX-S

Red, 0.4" high LCD

Resolution 0.1° below 375°F (190°C). Auto Ranges to 1˚ above 375°F (190°C).

Meter Accuracy at 77°F

±0.1°F/C ±1 digit at 32°F (0°C)

System Accuracy

(Meter with 202

immersion probe)

±0.1°F ±1 digit

at ice point,

±0.25% reading

thereafter

±0.3°F at ice point, 

±0.5% reading thereafter

Repeatability

±0.2°F

Peak Hold

Microcomputer calculated, switch to “P” to activate

Ambient Operating

Range

0 to 130°F (-18 to +55°C)

Temperature

Coefficient

0.01 deg/deg over ambient range

FM (Factory Mutual)

Approved*

Class I & II,

Div. 1, Groups:

D, F, G

Class I & II,

Div. 1, Groups:

C, D, E, F, G

Not 

Available

Power

11681-1

6.25 V

NiCad Battery

11681-2

6.25 V

NiCad Battery

9V Alkaline

battery 

(NEDA 1604A)

Battery Life

4.5 hrs. per

charge

392E/392E-S

10 hrs. per charge

50 hrs.

per charge

Low battery

indication.

392EX/392EX-S

50 hrs. per charge

Low Battery Indicator

Yes

Noise Rejection

Greater than 40 dB at 60 Hz increasing at 20 dB per decade.
Spurious noise attenuated by internal 16 Hz filter.
392HP includes special noise rejection program for use in presence
of rampant EMF such as spark ignition engines.

Size

3" W x 6" L x 1.5" D (7.5 cm x 15 cm x 3.75cm)

Weight

12 oz. (340gm)

7 oz. (198gm)

* FM Approval Groups: C (Ethylene Gas), (Methane, Propane), (Metal Dust), (Coal
Dust), and (Grain Dust)

All 392 Models Available in °F, °C, or Dual Scale, add “F”, “C” or “D” to Model Number. Dual
Scale not available on -S Models.

392 S

ERIES

M

ETER

S

PECIFICATIONS

Model

392M
392MX

392M-S
392MX-S

392VX

Range

-60 to 752°F

-51 to 400°C

-60 to 752°F

-51 to 400°C

Digital Display

392M/392M-S

Bright 0.33" high LED

Red, 0.4" high LCD

392MX/392MX-S

Red, 0.4" high LCD

Resolution 0.1° below 375°F (190°C). Auto Ranges to 1˚ above 375°F (190°C).

Meter Accuracy at 77°F

±0.1°F/C ±1 digit at 32°F (0°C)

System Accuracy

(Meter with 202

immersion probe)

±0.3°F at ice point, 

±0.5% reading thereafter

Repeatability

±0.2°F

Peak Hold

Microcomputer calculated, switch to “P” to activate

Ambient Operating

Range

0 to 130°F (-18 to +55°C)

Temperature

Coefficient

0.01 deg/deg over ambient range

FM (Factory Mutual)

Approved*

Class I & II,

Div. 1, Groups:

D, F, G

Class I & II,

Div. 1, Groups:

C, D, E, F, G

Not 

Available

Power

11681-1

6.25 V

NiCad Battery

11681-2

6.25 V

NiCad Battery

9V Alkaline battery

(NEDA 1604A)

Battery Life

392M/392M-S

10 hrs. per charge

50 hrs.

per charge

392MX/392MX-S

50 hrs. per charge

Low Battery Indicator

Yes

Noise Rejection

Greater than 40 dB at 60 Hz increasing at 20 dB per decade.
Spurious noise attenuated by internal 16 Hz filter.
392HP includes special noise rejection program for use in presence
of rampant EMF such as spark ignition engines.

Size

3" W x 6" L x 1.5" D (7.5 cm x 15 cm x 3.75cm)

Weight

12 oz. (340gm)

7 oz. (198gm)

* FM Approval Groups: C (Ethylene Gas), (Methane, Propane), (Metal Dust), (Coal
Dust), and (Grain Dust)

All 392 Models Available in °F, °C, or Dual Scale, add “F”, “C” or “D” to Model Number. Dual
Scale not available on -S Models.

4.0  SPECIFICATIONS

Содержание HEAT-Prober 392HP

Страница 1: ...rober PLATINUM RTD THERMOMETERS Models 392HP 392E 392E S 392EX 392EX S 392EVX 392M 392M S 392MX 392MX S 392VX 360X and 700M INSTRUCTION MANUAL Copyright 2004 Wahl Instruments Inc Printed in USA W 392...

Страница 2: ...authorized to assume for Wahl any liablility except as set forth above Warranty Temperature Probes and Meter Accessories Wahl warrants its Hand Held Meter Temperature Probes to be free of defects in...

Страница 3: ...in all light conditions To minimize power drain the display is multiplexed The LCD display is 0 4 high with excellent contrast A large portion of the circuit is designed with low power C MOS componen...

Страница 4: ...to conserve battery life 2 1 Meter Operation The HEAT Prober instrument has an accurate measuring range as stated on the front label The probes however may not be capable of this entire range of use...

Страница 5: ...ble of 50 hours continuous operation 2 4 1 Recharging To recharge the batteries use the recharger supplied Plug in directly to 115VAC 60 cycle wall plug and insert jack into socket on left side of HEA...

Страница 6: ...body temperature it requires good contact with that body or in the case of liquids or gases good immersion and mixing A portable probe temperature sensor must have the following characteristics a a l...

Страница 7: ...hane Propane E Metal Dust F Coal Dust and G Grain Dust All 392 Models Available in F C or Dual Scale add F C or D to Model Number Dual Scale not available on S Models 392 SERIES METER SPECIFICATIONS M...

Страница 8: ...gure 5 1 392M Platinum Heat Prober with Wahl Calibrator Model Number CAL392HT 6 0 SET UP AND CALIBRATION 6 1 Equipment Needed A Digital voltmeter Fluke 41 2 digit Model 8600A or equivalent B Wahl Plat...

Страница 9: ...a quick and accurate calibration can be achieved 6 6 1 Setup and Calculations The decade resistor box may be used as the source to perform a standard calibration A coiled cord cable with connector fo...

Страница 10: ...nd 392EVX Models 392HP 392M and 392E Use the Circuit Board Diagrams on these pages to locate potentiometer placement of Zero Adjustment and Span Adjustment Note that placement varies with Model See pa...

Страница 11: ...entire temperature range This also results in superior system accuracies at all temperatures The HP option utilizes a Noise Rejection program which minimizes transient noise effects In this version a...

Страница 12: ...he program Details of the components will be discussed in this section 9 2 Program Flow When Power is turned on the program will start executing instructions from address 000 the location of the top l...

Страница 13: ...age be crammed with new data Additionally if over range is displayed indicating on open probe lead or removed probe conditions are set for the next entry into Gaincheck to cause the compensation readi...

Страница 14: ...58MHz square wave with levels from 0 to 5 volts Replace U6 If still no display go to Step 9 Replace U4 If still no dis play go to Step 7 4 Check voltage at Q4 collector It should be equal to the batte...

Страница 15: ...Return unit to factory for repair NOTE Does not apply to 392MX and 392M with Serial Numbers before N120 11 0 ACCESSORIES 392 SERIES CALIBRATORS Model Number Usable With Meter Calibration Points CAL39...

Страница 16: ...the same as the 392M which is a recharge able nickel cadmium battery pack composed of five AA size cells with a com bined voltage of 6 0 volts and capacity rating of 65 amp hours The unit is supplied...

Страница 17: ...0 plug connector 3 Set resistance to 29 490 ohms and adjust R5 for a display read ing of 0 00 C 32 F 4 Set the decade box to a resistance of 2760 ohms and adjust R13 for a display reading of 60 00 C 1...

Страница 18: ...to next step 8 Measure voltage at pin 6 of wire harness It should be equal to the battery or power supply voltage Proceed to next step Check wire harness for broken wires 9 Replace Q1 and repeat Step...

Страница 19: ...follow up reminders about recalibration dates 16 0 RETURN FOR REPAIR CALIBRATION SERVICE Warranty Repair Information Please contact Wahl Customer Service at 800 421 2853 for a Return Merchandise Auth...

Отзывы: