12.0  CALIBRATION SERVICE

Recalibration service can be provided at a nominal cost, if desired. Special
thermal calibration traceable to NIST is available. Standard temperature
points are: 0°C, 94°C, and 250°C. Custom points are available upon request.

13.0  700M THERMISTOR .01°C/F RESOLUTION HEAT PROBER

13.1  General Discussion
The Wahl Thermistor unit measures temperature over a range of -1.00°C to
60.99°C (30.20°F to 141.78°F) with .01°C resolution. The sensing element in
the probes is a highly stable thermistor, with .1°C interchangeability, the out-
put being processed and linearized before display. The accuracy of the meter
is .02°C. With probe installed, the accuracy is .04°C.

The signal conditioning circuitry is completely chopper stabilized and associ-
ated resistive components have a very low temperature coefficient. This com-
bination yields essentially drift free operation.

The linearization of the Thermistor sensor is accomplished in two steps for
best utilization of the available analog-to-digital converter range. A resistive
network accomplishes a portion of the linearization while a look-up” table in
the microprocessor completes the balance.

All operations of the instrument are controlled by the microprocessor includ-
ing the drive for the display and peak reading computation and retention.
Another important feature of the instrument is the use of dual power regula-
tors and dual grounding systems to provide the ultimate in isolation between
sensitive signal conditioning circuits and the high level digital processing and
display circuits.

The power source of the 700M is the same as the 392M, which is a recharge-
able nickel-cadmium battery pack composed of five AA size cells with a com-
bined voltage of 6.0 volts and capacity rating of .65 amp-hours. The unit is
supplied with a recharger which can be plugged in at any time to maintain
the charge of the battery pack. Low battery condition is displayed as “L” in
the left-hand side of the display. The 700M is available only with LED display.

Should the temperature being measured extend beyond the limits of the
instrument, or in the event the probe is removed from the unit or becomes
defective and opens, an indication of -OL- will be displayed on the LED dis-
play. Correcting the over/under range condition or inserting a functioning
probe will again bring the instrument to a proper indicating condition.

T

OLERANCE

(P

ROBE

I

NTERCHANGEABILITY

)

Temp °C

-250

-220

-200

-100

0

100

200

300

400

500

Tolerance
±°C

5.8

1.8

1.2

0.7

0.3

0.5

1.1

1.7

2.3

3.0

±:

1.0

0.7

0.5

0.3

0.1

0.2

0.4

0.6

0.8

1.0

T

OLERANCE

(P

ROBE

I

NTERCHANGEABILITY

)

°F

0

-20

-40

-60

-80

-100

-120

-140

-160

-180

-200

0

93.03

88.65

84.26

79.85

75.42

70.98

66.52

64.04

57.54

°F

0

20

40

60

80

100

120

140

160

180

200

0

93.03

97.39 101.74 106.07 110.38 114.68 118.97 123.24 127.50 131.74 135.97

200

135.97 140.18 144.38 148.57 152.74 156.90 161.04 165.17 169.28 173.28 177.47

400

177.47 181.54 185.60 189.64 193.67 197.69 201.69 205.68 209.65 213.61 217.55

600

217.55 221.49 225.40 229.31 233.20 237.07 240.93 244.78 248.61 252.43 265.23

800

256.23 260.02 263.80 267.56 271.31 275.04 278.76 282.46 286.15 289.83 293.49

T

OLERANCE

(P

ROBE

I

NTERCHANGEABILITY

)

Temp
°C

-420 -400 -350 -300 -200 -100

0

32

100 200 300 400 500 600 700 800 900

Tolerance
±°C

10.4

8.3

3.3

1.9

1.5

1.1

0.7 0.6

0.7

0.9

1.5

2.0

2.7

3.3

3.9

4.5

5.4

±:

1.0

0.9

0.7

0.5

0.4

0.3

0.2 0.1

0.2

0.2

0.3

0.4

0.6

0.7

0.8

0.9

1.0

Содержание HEAT-Prober 392HP

Страница 1: ...rober PLATINUM RTD THERMOMETERS Models 392HP 392E 392E S 392EX 392EX S 392EVX 392M 392M S 392MX 392MX S 392VX 360X and 700M INSTRUCTION MANUAL Copyright 2004 Wahl Instruments Inc Printed in USA W 392...

Страница 2: ...authorized to assume for Wahl any liablility except as set forth above Warranty Temperature Probes and Meter Accessories Wahl warrants its Hand Held Meter Temperature Probes to be free of defects in...

Страница 3: ...in all light conditions To minimize power drain the display is multiplexed The LCD display is 0 4 high with excellent contrast A large portion of the circuit is designed with low power C MOS componen...

Страница 4: ...to conserve battery life 2 1 Meter Operation The HEAT Prober instrument has an accurate measuring range as stated on the front label The probes however may not be capable of this entire range of use...

Страница 5: ...ble of 50 hours continuous operation 2 4 1 Recharging To recharge the batteries use the recharger supplied Plug in directly to 115VAC 60 cycle wall plug and insert jack into socket on left side of HEA...

Страница 6: ...body temperature it requires good contact with that body or in the case of liquids or gases good immersion and mixing A portable probe temperature sensor must have the following characteristics a a l...

Страница 7: ...hane Propane E Metal Dust F Coal Dust and G Grain Dust All 392 Models Available in F C or Dual Scale add F C or D to Model Number Dual Scale not available on S Models 392 SERIES METER SPECIFICATIONS M...

Страница 8: ...gure 5 1 392M Platinum Heat Prober with Wahl Calibrator Model Number CAL392HT 6 0 SET UP AND CALIBRATION 6 1 Equipment Needed A Digital voltmeter Fluke 41 2 digit Model 8600A or equivalent B Wahl Plat...

Страница 9: ...a quick and accurate calibration can be achieved 6 6 1 Setup and Calculations The decade resistor box may be used as the source to perform a standard calibration A coiled cord cable with connector fo...

Страница 10: ...nd 392EVX Models 392HP 392M and 392E Use the Circuit Board Diagrams on these pages to locate potentiometer placement of Zero Adjustment and Span Adjustment Note that placement varies with Model See pa...

Страница 11: ...entire temperature range This also results in superior system accuracies at all temperatures The HP option utilizes a Noise Rejection program which minimizes transient noise effects In this version a...

Страница 12: ...he program Details of the components will be discussed in this section 9 2 Program Flow When Power is turned on the program will start executing instructions from address 000 the location of the top l...

Страница 13: ...age be crammed with new data Additionally if over range is displayed indicating on open probe lead or removed probe conditions are set for the next entry into Gaincheck to cause the compensation readi...

Страница 14: ...58MHz square wave with levels from 0 to 5 volts Replace U6 If still no display go to Step 9 Replace U4 If still no dis play go to Step 7 4 Check voltage at Q4 collector It should be equal to the batte...

Страница 15: ...Return unit to factory for repair NOTE Does not apply to 392MX and 392M with Serial Numbers before N120 11 0 ACCESSORIES 392 SERIES CALIBRATORS Model Number Usable With Meter Calibration Points CAL39...

Страница 16: ...the same as the 392M which is a recharge able nickel cadmium battery pack composed of five AA size cells with a com bined voltage of 6 0 volts and capacity rating of 65 amp hours The unit is supplied...

Страница 17: ...0 plug connector 3 Set resistance to 29 490 ohms and adjust R5 for a display read ing of 0 00 C 32 F 4 Set the decade box to a resistance of 2760 ohms and adjust R13 for a display reading of 60 00 C 1...

Страница 18: ...to next step 8 Measure voltage at pin 6 of wire harness It should be equal to the battery or power supply voltage Proceed to next step Check wire harness for broken wires 9 Replace Q1 and repeat Step...

Страница 19: ...follow up reminders about recalibration dates 16 0 RETURN FOR REPAIR CALIBRATION SERVICE Warranty Repair Information Please contact Wahl Customer Service at 800 421 2853 for a Return Merchandise Auth...

Отзывы: