![Wagner Meters Orion 910 Скачать руководство пользователя страница 15](http://html.mh-extra.com/html/wagner-meters/orion-910/orion-910_instruction-manual_868301015.webp)
25
26
Turn the meter on, and then use the
SPECIES/MATERIAL button to put your
meter in CAL mode. (Refer to the SPECIES/
MATERIAL button section of the manual.)
Place the meter on the CR, correctly
“seating” the meter sensor pad firmly on
the recessed area of the CR. IMPORTANT:
Failure to correctly and firmly “seat” the
sensor pad in the recessed area will cause an
inaccurate calibration.
Apply light downward pressure and press
the AUDIO button three (3) times. The
meter will make a “beeping” sound which
indicates it is going through the automatic
calibration procedure.
When the calibration procedure is
complete, the meter will display the word
LIFT. Immediately remove the meter from
the CR and hold the meter in the air for
approximately 5 seconds until the word
DONE appears on the display. During
this last phase in the air, be sure to keep
your hand and other objects away from the
underside of the meter.
Press the ON/HOLD button to return the
meter to Standard Measuring mode.
To restore your meter to the original factory
default settings, including the original
calibration, press and hold the ON/HOLD
button for approximately 10 seconds, until
the word CAL appears on the display, then
release the ON/HOLD button.
If, after recalibrating the Orion® 910,
there is still a problem with the meter,
please contact a Wagner representative for
further instructions at 844-689-0660.
Protective Rubber Boot
for Your Orion® 910
In an effort to help avoid
damage resulting from
your Orion® 910 meter
being dropped or banging
the edges against other hard
objects, Wagner Meters
has provided a fitted rubber boot which
fits snugly around the perimeter of your
Orion® 910 meter.