Calibration of Your Orion® 910
If you have reason to believe your Orion®
910 may be out of calibration, it can be easily
recalibrated using the provided Calibration
Reference (CR). Before calibrating your
meter, please be aware that you must
ONLY calibrate to the correct CR that came
with the meter. Serial numbers on the CR
(found underneath) and meter (found in
the battery compartment) MUST match for
correct calibration. Also, the sticker must
not be damaged, as this may cause incorrect
calibration. Follow the steps below for
proper calibration. The instructions are also
printed on the back of the CR. Place the CR
on its legs on a non-metallic surface.
Low Battery Alert Indication
When the battery is low, BAT appears in
the upper right-hand corner of the display.
If ignored for too long, the meter will
flash a larger BAT three times and then
automatically shut down. The 9-volt battery
must be replaced immediately.
This shut-down action is to prevent
inaccurate readings.
Auto Shut-Down to Preserve
Battery Life
Regardless of which mode the meter is
in, if there is no change in measurement
activity over a period of 1 minute, the
meter will automatically turn off.