User manual - DMS9648O
87
8
Offset calibration
Analogue input offset correction. Value added / subtracted to the process
visualization.
-10000..+10000
[digit
1
],
Default
0.0.
9
Gain calibration
Percentage value that is multiplied for the process value (allows to
calibrated the working point)
-100.0%..+100.0%,
Default
: 0.0
ex: to correct the range from 0..1000kg showing 0..1010kg, set the
parameter to -1.0.
10 Calibration
Selection of calibration type for the analogue input. After finishing or
deleting calibration proceeding, this parameter will automatically return
to default setting (=Disabled).
Disabled (
Default
)
Full scale % value
Sampling value
mV/V Value
11 Totalizer
Visualize on corresponding page the total value of measured process
considering the sensor signal as unit/time value (ex. if full scale of the
connected sensor is 2000m³/hour, parameter 11
Totalizer
has to be
selected as
Hour
and display will visualize the total fluid volume from the
last RESET/START signal till present).
Disabled
Display visualizes process value (
Default
)
Second
Display visualizes rate in unit/s
Minute
Display visualizes rate in unit/min
Hour
Display visualizes rate in unit/hour